background image

General Operations Cont.

General Operations Cont.

Default display and operation procedure after unit power up

 

Default display and operation procedure after unit power up cont.

 

Default display and operation procedure after unit power up cont.

 

Default display and operation procedure after unit power up cont.

 

Default display and operation procedure after unit power up cont.

 

5.

 Press both buttons 

“UP”+ ”Down”

 at the same time for 

Celsius and Fahrenheit unit selection. Default is ºF.

6. 

There is one three-position slide switch inside the battery 

compartment. The user can enable the probe by selecting 
the position of the switch. When the switch is in the “

L

” 

position, the left probe is working and the right sensor 
buttons are functionless. When the switch is in the “

R

” 

position, the right probe is working and the left sensor 
buttons are functionless. When the switch is in the middle 
position, both probes are working and all buttons are 
functional. The probe should be plugged into the unit all 
the time as long as it is enabled. Position of the slide switch 
should be in the correct position before battery is installed. 
Changing the slide switch position requires resetting the 
unit by re-installation of battery or pressing the system reset 
button, which is also inside the battery compartment.

7.

 If the probe is enabled (through the slide switch) and is 
unplugged during unit powering up or setup, the alarm will 
beep and EEE will be displayed in the corresponding probe 
temp display area.  The unit cannot be set up until the 
probe is plugged in again.

8.

 During normal operation, when the probe is unplugged, 
the EEE temp and duration of the probe being unplugged 
will be recorded. The alarm will beep during the probe is 
unplugged.  After the probed is inserted again, the 
operation for that probe resumes.

9.

 During the time the probe is plugged-in or unplugged, the 
temperatures recorded at that moment may not be 
accurate.  If these temperatures meet the temp high/low 
settings, they will be written to the alarm records (together 
with the “EEE” alarm record), and the users need to filter 
out these un-accurate temperature records themselves.

10.

 The

 "Password Reset"

 button is inside the battery 

compartment.  Pressing the button once will reset the 
password to 

“88:88”

.  

11.

The 

“System Reset”

 button is inside the battery 

compartment.  The user can reset the unit any time for re-
initialization of the unit or

 

in case the unit does not start up 

after installation of the batteries. After this button is pressed, 
the unit will go through the normal power up initialization 
process, and all settings have to be re-entered.  Only the old 
alarm records (if there is any) can be retrieved after system 
reset.  The operator may use a paper clip or similar type of 
object for accessing the system reset button. 

    

Note: Old alarm records may be retrieved after hitting 

System Reset button, but when alarm settings are re-
entered old records will be deleted.

1. 

All LCD segments are displayed for 3 seconds and the alarm 

sounds for 1 second.

2. 

Then, current temperature(s) is displayed (default is °F), left 

hand side of the bottom line displays 

“88:88”

, right hand 

side of the bottom line displays a flashing 

“00H00M00S”

.

3. 

The unit enters the clock setting mode (flashing 

00H00M00S). Press 

“UP/DOWN”

 buttons for Hour setting. 

Pressing down the

 “UP/DOWN”

 buttons for more than 3 

seconds activates the fast forward function.  Press the 

“SET”

 

button to confirm the Hour setting.  Hour digits stop flashing 
and the remaining Minute and Second digits continue to 
flash.  Press 

“UP/DOWN”

 buttons for Minute setting.  

Pressing down the

 “UP/DOWN”

 buttons for more than 3 

seconds activates the fast forward function. Press the 

“SET”

 

button to confirm the Minute setting. Hour and minute digits 
stop flashing and the remaining Second digits continue to 
flash.  Press 

“UP/DOWN”

 buttons for Second setting. 

Pressing down the 

“UP/DOWN”

 buttons for more than 3 

seconds activates the fast forward function. Press the 

“SET”

 

button to confirm the Second setting. The clock digits stop 
flashing and start to tick. Left or right or both sides buttons 
can be used for clock setting depending or which probe(s) is 
enabled.

4.

 After the clock is set, the unit enters the password setting 
mode (

flashing “88:88”

).  Left or right or both sides buttons 

can be used for setting depending or which probe(s) are 
enabled.

(I)

 Press the 

“Up”

 or 

“Down”

 button for selecting the number 

from 0~9 for first digit (the first press of 

“Down”

 changes 

the digit from 8 to 7; the first press of up changes the digit 
from 8 to 9).  For example, 5 is selected. Pressing the 

“Set”

 

button again confirms 5, and the other three segments will 
remain flashing.

(ii)

 Press the 

“Up”

 or 

“Down”

 button for selecting the number 

from 0~9 (ex.4 is to be selected), press the 

“Set”

 button 

again confirms 4, and the other two segments will remain 
flashing.

(iii)

 Press the 

“Up”

 or “

Down” 

button for selecting the number 

from 0~9 (ex.3 is to be selected), press the 

“Set”

 button 

again the 3 is confirmed and the last 8 still flashing.

(iv)

 Press the 

“Up”

 or 

“Down”

 button for selecting the number 

from 0~9 (ex.2 is to be selected),press the

 “Set”

 button 

again the 2 is confirmed and the Password 

“54:32”

 is set. 

(v)

 The password display will show the set status of the probes: 

“Set0”

 meaning that none of the probes is set. After the 

password is set, the user can lock the unit by pressing the 

“PASSWORD”

 button once. When the unit is locked, the 

“LOCK”

 icon on the display will be turned on to indicate that 

password has to be re-entered again for accessing the unit. 

(vi)

 After the unit is locked, the user has to enter the password 

to access the unit. Press the 

“PASSWORD”

 once, flashing

 

“88:88”

 appears on the left hand side of the bottom line. 

Enter the password by following the set password procedure. 
If the password is correct, the password will stop blinking 
after the last digit entry and the display will revert to 

“SetL/R/2/0”

. All buttons are available at this time. 

Otherwise, the password will just keep blinking and the 
password entry process has to re-start again from the first 
password digit.

(vii)

 The default password of the unit is 

“88:88”

, the first 

change of the password will be recorded as the new 
password. In case the password is forgotten, the user has to 
open the battery door and press the 

“Password Reset”

 

button on the battery compartment and the password will be 
restored to the default password of 

“88:88”

. The user can 

then access the unit with the default password 

“88:88”

. To 

set a new password, the user has to reset the unit or re-
install the battery.  The unit will start the whole setup process 
after system reset or battery re-installation.  All previous 
settings will be lost except that the old alarm event records 
are still retrievable.  The high / low alarm set points and 
delays are reset to the default values.  The alarm 
temperatures and/or the delay time should not be set for old 
alarm records retrieval as new alarm settings will clear the 
old alarm memory contents. The old alarm records will only 
be cleared after the alarm temperatures and delay setting is 
complete. The unit will quit the alarm setting mode 
automatically during the alarm setting process if there is no 
button pressed for 1 minute.  This way the old alarm records 
are preserved.

5. 

Once the clock and password are set, the unit enters the 

normal display mode. The status of the probe setting is 
displayed on the left hand bottom line of the LCD as 

“Set0”

“Setr”

“SetL”

, or 

“Set2”

. The user can enter the set high / 

low alarm events mode or scroll alarm events mode from 
the normal display mode. If the user does not set the high / 
low alarm set points, the unit will not record any alarm 
events and will just display the current temperatures.

(i)

 

Set High/Low Alarm set point:

a. 

The default high/low alarm set points are 40.0°F and 32.0°F 

respectively while the default delay time is 00H00M00S for 
both high and low alarms.

b. 

To set the alarm set points and corresponding delay times, 

press down the 

“SET”

 button on Left or Right side for 3 

seconds, the corresponding temperature display will display 
the default high (40.0°F) or previous high alarm set point 

    The digit behind the decimal displays 

“H”

 and the left hand 

side of the bottom line displays 

“Set”

 and a flashing 

“R”

 or

 

“L”

 indicating that the right or left probe is being set. 

Increase or decrease the high alarm set point by pressing 

the 

“UP” 

or

 “DOWN”

 key. When the 

“UP” 

or

 “DOWN”

 

key is pressed for more than 3 seconds, the fast forward 

function is activated. When the desired high alarm set point 

is reached, press the 

“SET”

 button to confirm high alarm set 

point and the high alarm set point stops flashing. Then 

“00H00M00S”

 or the previous high alarm delay starts 

flashing on the right hand side of the bottom line for delay 

time setting. Press 

“UP/DOWN”

 buttons for Hour setting. 

Pressing down the 

“UP/DOWN”

 buttons for more than 3 

seconds activates the fast forward function. Press the 

“SET”

 

button to confirm the Hour setting. Hour digits stop flashing 

and he remaining Minute and Second digits continue to 

flash. Press 

“UP/DOWN”

 buttons for Minute setting. 

Pressing down the 

“UP/DOWN”

 buttons for more than 3 

seconds activates the fast forward function. Press the 

“SET”

 

button to confirm the Minute setting. Hour and minute digits 

stop flashing and the remaining Second digits continue to 

flash. Press 

“UP/DOWN”

 buttons for Second setting. 

Pressing down the 

“UP/DOWN”

 buttons for more than 3 

seconds activates the fast forward function. Press the 

“SET”

 

button to confirm the Second setting. The alarm delay time 

stops flashing. At this point, the high alarm set point and 

delay are set.

c.

 The temperature display then displays the flashing default 

(32.0°F) or previous low alarm set point, and the digit 

behind the decimal will display 

“L”

.  Increase or decrease 

the value by pressing the 

“UP

” or 

“DOWN” 

button.  When 

the 

“UP”

 or 

“DOWN”

 button is pressed down for more 

than 3 seconds, the fast forward function is activated.  When 

the desired low alarm set point is reached, press the 

“SET”

 

button again to confirm the low alarm set point and the low 

alarm set point stops flashing.  (Note: the low alarm set 

point must be smaller than the high alarm set point.  

Otherwise the low alarm set point will keep flashing after 

the 

“SET”

 button is pressed for low alarm set point 

confirmation.)  Then 

“00H00M00S”

 or the previous low 

alarm delay starts flashing on the right hand side of the

    bottom line for low alarm delay setting. Press 

“UP/DOWN”

 

buttons for Hour setting.  Pressing down the 

“UP/DOWN”

 

buttons for more than 3 seconds activates the fast forward 

function.  Press the 

“SET”

 button to confirm the Hour 

setting. Hour digits stop flashing and the remaining Minute 

and Second digits continue to flash. Press 

“UP/DOWN”

 

buttons for Minute setting. Pressing down the 

“UP/DOWN”

 

buttons for more than 3 seconds activates the fast forward 

function. Press the 

“SET”

 button to confirm the Minute 

setting. Hour and Minute digits stop flashing and the 

remaining Second digits continue to flash. Press 

“UP/DOWN”

 buttons for Second setting. Pressing down the

 

“UP/DOWN”

 buttons for more than 3 seconds activates the 

fast forward function. Press the 

“SET”

 button to confirm the 

Second setting. The alarm delay time stops flashing. At this 

point, the low alarm and its delay time are set.

d.

 Once the high / low alarm and delays are set, the left hand 

side of the bottom line displays 

“Set” “R” 

or

 “L”

 indicating 

that the right or left probe is already set. The display then 

returns to the normal display of current temperature, the 

clock, and 

“Set” “R” 

or

 “L” or “2”

.

e.

 Repeat the above setting procedure for another probe when 

appropriate. If both probes are set, the normal display 

should show the current temperature of both probes, the 

clock, and “Set2”.

When the temperature is over the high point or below the low 
point, the alarming events are recorded. There are maximum 
96 Hours alarm history stored in memory. In the normal display 
mode, press the 

“SET”

 button once to display the last alarm 

event.When the current displayed alarm event is the last one,

 

pressing 

“UP”

” 

will scroll the alarm event back to the oldest

 

alarm event

 For example, if the oldest alarm event is D0 R: 

14:15 A 40°F (high temperature alarm occurred on the current 
day at 14:15). Pressing the 

“UP”

 button once displays D0 R: 

14:16 44°F (the maximum temperature recorded during alarm 
period was 44°F at 14:16).

Pressing the 

“UP”

 button again will display D0 R: 14:18 

00h02m33s (the temperature went back into safe range at 

14:18, and the duration of the alarm was 2 minutes, 33 

seconds). Pressing the 

“UP”

 button again will move to next 

record.Pressing and holding the 

“UP”

 button for 3 seconds 

activates the fast forward function. The alarm events are scrolled 

backward by pressing the 

“DOWN”

 button once.  Pressing and 

holding the 

“DOWN”

 button for 3 seconds activates the fast 

backward function. The last displayed alarm record will stay on 

the display for 15 seconds.  After 15 seconds the normal display 

(current temperature, 

“Set” “R/L/2/0”

, Clock) is resumed. 

Once in the alarm events display mode, the only way to exit to 

the normal display mode is to wait for the 15 seconds alarm 

event time out or when a new alarm event arrives. The 

temperature display for the corresponding probe should be 

blank or only displaying the temperature data associated with 

the alarm record when in alarm events display mode.  The set 

high / low alarm function is disabled in the alarm events display 

mode, i.e. even though the user presses the 

“SET”

 for 3 

seconds, the unit will not enter the alarm events set mode.
When there is no alarm records, “

nr

” (no records) will be 

displayed on the lower left line of display when the 

“SET”

 button 

is pressed (clock is still displayed on the lower right line).  

Normal display mode is resumed after the 10 seconds alarm 

events display time-out time. 
The alarm records can't be deleted, unless new alarming set 

points are set.  It always keeps maximum 96 hours records in the 

memory.  The oldest alarm record will be replaced by the new 

one when the memory is full.
When the new high and/or low alarm value is set, the memory 

will clear all the records and start to record the new alarming 

date from zero.
There are total 120 alarm records, 60 alarm records per probe.

Example of alarm records: 
Initial settings: 

High alarm = 40.0°F

 

Low alarm = 32.0°F
High alarm delay = 20 minutes

 

Low alarm delay = 30 minutes

D3HL 6:35  39.1°F (high of the day was 39.1°F at 6:35:20am, 39.1°F is displayed in temperature 

display screen)
D3LL 17:22  37.0°F (low of the day was 37.0°F at 5:22:45pm, 37.0°F is displayed in temperature 

display screen)
D2HL 5:15  38.6°F (high of the day was 38.6°F at 5:15:00 am, 38.6°F is displayed in temperature 

display screen)
D2LL 12:01  33.0°F (low of the was 33.0°F at 12:01:30pm, 33.0°F is displayed in temperature 

display screen)
D1 L 14:00 A 40.0°F (at 2:00:15 pm temperature went above 40.0°F, 40.0°F is displayed in 

temperature display screen) 
At 14:20 the audible alarm starts beeping.
D1 L 14:30  50.0°F (high temperature for the alarm is 50.0°F at 2:30:30pm, 50.0°F is displayed in 

temperature display screen)
D1 L 14:40 00h40m15s (temp went back below 40°F at 2:40:30pm after 40minutes and 15 

seconds of being out)
D1 L 14:45 Usr (user stops alarm at 2:45:55pm and changes temperature unit from °F to °C)
D1 L 18:00 b0.0°C (at 6:00:23pm temperature went below 32.0°F, 0.0°C is displayed in 

temperature display screen)
D1 L 18:05 –0.6°C (low temp for the alarm is 30.9°F at 6:05:01 pm, -0.6°C is displayed in 

temperature display screen)
At 18:30 the audible alarm starts beeping.
D1 L 20:00 US r (user stops alarm at 8:00:59pm and changes temperature unit from °C to °F)
D1 L 20:10 2h09m42s (temp went back above 32°F at 8:10:05pm after 2 hours and 9 minutes 42 

seconds of being out)
D1 L 21:00 b32°F (at 9:00:00pm temp went below 32 degrees, 32°F is displayed in temperature 

display screen)
D1 L 21:01 29.0°F (the low temp for the alarm is 29.0°F recorded at 9:01:08pm, 29.0°F is 

displayed in temperature display screen) 
D1 L 21:02 00h02m30s (Temp went back above 32.0°F at 9:02:30pm after 2 minutes and 30 

seconds of being out)
D1HL 14:30 50.0°F (high of the day was 50 degrees at 2:30:30pm, 50°F is displayed in 

temperature display screen)
D1LL 21:01 29.0°F (low of the day was 29 degrees at 9:01pm, 29.0°F is displayed in temperature 

display screen)
Between "D1 L 14:00 A40.0°F" & "D1 L 14:30 50.0°F" I would put a note saying that at 14:20 the 

audible alarm starts beeping.
Between "D1 L 18:05 -0.6°C" & "D1 L 20:00 US r" I would put a note saying that at 18:30 the 

audible alarm starts beeping.
Between "D0 L 10:00 A40.0°F" & "D0 L 10:30 45.3°F" I would put a note saying that at 10:20 the 

audible alarm starts beeping.
D0 L 10:00 A40.0°F (at 10:00:05 am temperature went above 40 degrees, 40.0°F is displayed in 

temperature display screen)
At 10:20 the audible alarm starts beeping.
D0 L 10:30 45.3°F (high temperature for the alarm is 45 at 10:30:01am, 45.3°F is displayed in 

temperature display screen)
D0 L 11:40 1h40m43s (temp went back below 40 at 11:40:48am after 1 hour 40minutes 43 

seconds of being out)
D0 L 12:00 b32.0°F (at 12:00:00 am temperature went below 32.0°F, 32.0°F is displayed in 

temperature display screen)
D0 L 12:00 US r (user stops alarm at 12:00:59pm) 

Note: Unit of temperatures records displayed is the current selected temperature unit 

regardless of the temperature unit when the record was taken.  For example, the unit of 

temperature was °C when the unit recorded an alarm event, and that alarm event is 

displayed in °F when the current temperature unit is °F and the alarm record is being 

scrolled back.

Alarm Records

Alarm Records Cont.

Alarm Record Format

Example Alarm Record Format:

Отзывы: