TAU STERIL
N° 11 - Rev. 3 03/02/20
User Manual
Page 24
FAST B
No
Message displayed
Description
Solution
33
DOOR MOTOR OR OPEN
MICROCONTACT
Door motor or close micro-switch
Contact technical assistance.
34
TRIAC Q13 OPEN SUPPLY
SHUT OFF
Electronic component
Turn the machine off and on again. If the problem per-
sist, contact technical assistance.
35
OK TO CLOSE DOOR AT
OFF
Door close authorisation/enable micro-
switch at OFF during cycle
Contact technical assistance.
36
DOOR MICROCONTACT
Door open micro-switch at ON during cy-
cle
Contact technical assistance.
37
DOOR NOT CLOSED DU-
RING CYCLE
Door closed micro-switch at OFF during
cycle
Contact technical assistance.
38
TEMP. R. CA. HIGH
High temperature resistance chamber
Contact technical assistance.
39
STEAM GEN. HEATER
TEMPERATURE HIGH
High temperature resistance generator
Contact technical assistance.
40
STARTING TO MUCH WA-
TER PUMP
Too many switching water pump
Contact technical assistance.
41
DOOR OPEN
Door closed not correctly
Contact technical assistance.
42
NO CALIBRATION E2P
Calibration data loss
Contact technical assistance.
43
CALIBRATION NON OPTI-
MAL
Reading incorrect calibration
Contact technical assistance.
4.9 Printer
The internal (see fig. 2) or external (see fig. 3) printer is used to print out the data for the complete sterilisation cycle
in order to guarantee a paper report to be included with the sterilised equipment.
To change
the time, date etc., see section 4.3.2, 4.3.3. To enable the internal printer, see section 4.3.5. For the exter-
nal printer, if this is an ordinary printer with parallel port connection, to install, proceed as follows:
Connect the cable to the 24-pin connector on the rear panel (see fig. 2).
Connect the
other end of the cable to the external printer.
Connect the printer supply using an appropriate cable.
The printer is now ready to print.
Note! If your internal printer is enabled, the external one will also work.
Example of paper strip printout showing the results of the sterilisation cycle (Fig. 10).
Fig.10