User Manual
Page 21
FAST B
4.7.1
Anti-condensation Ventilation Phase ( VEN )
Ventilation phase VEN: at the end of the sterilisation cycle, the display will show VEN (if enabled): anti-condensation
ventilation. The sterilisation chamber of the autoclave is vented every 20 sec. by means of a vacuum pump, until the
temperature shown on the display (T. Cham.) falls to 102° C.
This ventilation ensures that the equipment is perfectly dry.
The machine can, however, be opened at any time (without waiting for the end of the VEN phase) if you are in a hurry:
press the
“
stop
”
button and the machine will show the door open icon.
VEN phase
Autoclave in phase
DV drying
Autoclave in
phase VEN
If you press Stop
you will see ...
… the final screen page
with the door open icon.
End of cycle
It is possible that, at the end of sterilisation cycle (after the VEN phase), the display indicates alarm No. 11 “Cooling
Pause”
"
when the door is opened. The purpose of this is to allow (once the cycle is complete) the parts of the machine
subject to thermal stress (during the sterilisation cycle) to cool down.
4.7.2
Info Button
The machine memorises the last sterilisation cycle. By pressing the info button when it is shown on the screen, you
can access the archived data:
Info button
Data summary
General cycle
Phase by phase
summary
If there is a problem with the machine, these archives are useful, in particular, to technical assistance.