1)
Reference Image 5.6 on page 9
.
2) Connect the cooling plate to power source. Turn on the power switch
located on the rear of the machine. Green light will illuminate.
3) Press
SET
key; the compressor will stop running.
4) Press
(+)
or
(—)
to adjust the temperature.
5) Press the
SET
key to enter running mode.
* Turn on 20 minutes prior to use to allow for temperature to be achieved.
• There are two indicator lights beside the screen. If the indicator light
underneath is on, this means the compressor is working. If the indicator
light on top is on, this means the compressor has stopped working.
• If the actual temperature is lower than setting temperature, the compressor
will automatically stop running. When the actual temperature is higher than
setting temperature, the compressor will restart.
• For convenience, the preset temperature range is -5° to -10°C.
14 Tanner Scientific® TN1500 - Embedding Center
6.6 | Cooling Plate Operation