12. Specifications
13. Troubleshooting
Height
Hm
Hf
xx.x
xxx.x
xx
{ 0
˜0
˜1
Output data
Name of item
Header
Weight
BMI
Checksum
Output data (ASCII code)
Pattern
BMI mode
2 – 6 bytes
3 – 5 bytes
Control data
Control data
Control data
Model
Control data
Fix to 16
Fix to 1
Fix to 1
Fix to 1
2 Bytes fixed
1 Byte fixed
1 Byte fixed
1 Byte fixed
MO
"WB-3000"
9 Byte fixed
2 Bytes fixed
˜2
Wk
Wp
CS
xxx.x
3 – 5 bytes
MI
Note:
•The data are divided with commas (,) for each data.
•The terminator (end of the data) is CR (ASCII code 0DH), LF (ASCII code 0AH).
•The control data for the items 1 – 4 are for expansion. They are not used presently so the receiving side can ignore them.
•When the measurement unit is select the "lb",all the hight data is converted into “ft-in” and it is output.
•Because of its specifications, the scale will output an irrelevant signal approximately 0.1 second after data output. Make sure you ignore this
signal and do not import it.
[Output example (in the case to lb unit)]
Weight mode
kg mode
lb mode
Model
WB-3000
Strain Gauge Load Cell
440 lb / 0.2lb
0 - 440lb
0 - 200kg
22 - 440lb
10 - 200kg
Mechanical Height Rod
2ft 1in – 7ft
2ft – 8ft 2in 0.5in increment
440 lb / 0.2lb
2ft – 8ft 2in 0.5in increment
0.1 increments
20.8 x 15.0 x 54.4 in
15.0 x 15.0 x 2.0 in
Upper 5Digits and Lower 4Digits LCD
Height of numerals 1in / 24mm
RS-232C (D sub 9-pin Male connector) USB (B-type)
AC adapter (included) Center Minus
0.3 W max.
32°F - 95°F
25.3lb (without batteries)
DC 6V 200mA (LR6 - AA Alkaline Battery x 4 included)
Approximately 100hours of continuous use when using LR6
(AA Alkaline battery)
0°C - 35°C
11.5kg (without batteries)
528 x 380 x 1381 mm
380 x 380 x 50 mm
61cm- 250cm 1cm increments
200kg / 0.1kg
61cm- 250cm 1cm increments
64cm – 214cm
200kg / 0.1kg
Weight Measurement
Measurement System
Maximum Capacity / Minimum Graduation
Height Measurement
Range of Weight
Standard
Step on
Measurement System
Range of Height
Input Items
Height
Output Items
Display
Weight
Height
BMI
Size
Overall
Platform
Display
Output Data Interface
Power Source
Power Consumption
Temperature Range of usage
Weight of Equipment
Rated Power
Battery Life
If you are concerned that the scale may not be functioning correctly, please check the following point before requesting repairs.
Problem
Check Points
Nothing is displayed when
key
is pressed.
“Lo” is displayed.
Measurement is not accurate.
The weight does not stabilize
“
OL
”
is displayed
“
-OL
”
is displayed
“
Error
”
is displayed
•Check the connection of the AC adapter jack to the DC jack, and the connection of the AC
adapter plug to the wall outlet.
•The batteries may be weak. Try an AC adapter or new LR6 (AA) batteries.
•Check the terminals of the batteries for contamination
The batteries are running out. Please replace them immediately or use the AC adapter.
•Check that all feet of scale are stable and on a hard flat surface, carpeting depth should be
kept to a minimum.
•After having stepped down from the scale and pressed the ON key, perform the measure-
ment again.
•Accurate measurement may be impossible if the product is used where there is excessive
vibration.
•Try using the product in a different place.
•Are you moving during measurement?
•Is the measured weight within the range of weight measurement?
•Are you touching the height rod arm, switch, or another part during measurement?
•The range of weight (440lb/200kg) is exceeded.
•Was anything placed on the scale before measurement?
•Check whether the cable between the platform and display is disconnected, cut, or pinched.