Calibration

Note:

Your scale has been accurately calibrated at the

factory; however, there is always a possibility that
calibration has changed due to jarring in shipping. The
following procedure should be followed to check and/or
correct any error in calibration.

1. Depress On/Off switch on scale and wait for display 

to show “0.”

2. Place an accurately known weight on the platform.

3. If display is other than know weight, then scale is out

of calibration.

4. Remove the 4 screws nearest the face switch panel (2

on each side – near the aluminum trim). Slightly loosen
the remaining 4 screws (2 on each side), 

but do not

remove

. Lift off the top portion of the head.

Be careful,

since it is attached to the bottom portion with a
ribbon cable.

Looking at the bottom from of the head

(underneath), you will see 2 small adjustment screws.
The left metal screw (VR2) is for coarse adjustments; it
adjusts for large errors. The right plastic screw (VR1) is
for fine adjustments; it adjusts small errors.

5. Turn the left metal coarse calibration screw (VR2) in

small increments, 

CLOCKWISE to increase weight or

COUNTERCLOCKWISE to decrease weight.

In most

cases, this is the only adjustment needed.

6. If necessary, make adjustments with the right screw

(fine adjustment).

CLOCKWISE to decrease weight;

COUNTERCLOCKWISE to increase weight.

7. Remove weight and depress the Zero switch. Wait for a

display reading of zero. Again add known weight to
platform. If reading is correct, no further adjustments
are necessary. If reading is incorrect, repeat steps 5
and 6 until correct reading is obtained.

TROUBLESHOOTING

The following guide will help you solve common problems:

Taking a Weight Measurement

1. Once an object’s (blanket, etc.) weight has been 

zeroed out and the display is reading “0,”
have the patient stand on the scale platform.

2. The moving dashes will lock in and display the

patient’s actual weight.

3. Weight measurement should automatically be

displayed on the control panel box.

•  Measurement automatically defaults to pounds. If

kilograms are required, depress the Kilos switch.

•  If a reweigh is required, depress the Zero switch.

4. Excessive movement may prevent a repeatable

reading. In this case, a series of moving dashes
continues to appear on the console, rather than a
weight reading. If moving dashes persist, activate the
“Average Lock” by depressing the appropriate Pounds
or Kilos switch to obtain an average reading of the
patient’s weight.

Note:

This reading is now an instantaneous average 

of weights that the computer calculates.

5. After weight is obtained and recorded, have the patient

step off the scale.

6. Depress On/Off switch to turn scale off.

IMPORTANT:

Make certain the platform height is

adjusted so the wheels and platform bottom edge do 
not touch the floor or carpet. An incorrect patient weight
reading will result if bottom of the platform touches. (See
“Leveling the Platform” on page 3.)

CARE AND MAINTENANCE

Proper care and cleaning of all Tanita Corporation scales
and accessories is essential to ensure a long life of
accurate and effective operation. In addition to routine
cleaning, include a regular check of all structural
apparatus to ensure stability.

Cleaning

Disconnect battery charger from scale before cleaning.

1. All external surfaces can be wiped clean with a damp

cloth or tissue. Mild soap and water solution may be
used. Wipe unit dry with a clean, soft cloth.

2.

DO NOT 

immerse control panel box into cleaning or

other liquid solution.

3.

DO NOT

use isopropyl alcohol or other solutions to

clean surface of control panel box.

4. Check for overall appearance of the whole scale for

any obvious damage, wear and tear. Inspect the AC
charger for cord cracking or fraying or for broken or
bent prongs.

5

PROBLEM

Scale won’t turn on

Incorrect patient weight
readings

Display flashes “Lo-b”

Patient weight readings
consistently too high or
too low

SOLUTION

A. Make sure plug in firmly plugged into wall power

source.

B. Check fuses or circuit breaker to ensure voltage

is present in wall power source.

A. Platform too low and hitting platform feet or

carpet. Increase platform height. See “Leveling
the Platform” on page 3.

B. Person, other than person being weighed,

touching or leaning on scale handlebar.

C. Weight of wheelchair or chair not taken into

account. See “Operating Instructions” on page 4.

D. Platform not level. See “Leveling the Platform” on

page 3.

Recharge battery. See “Charging the Battery”
on page 3.

Check scale calibration and calibrate if necessary.
See “Calibration” on page 5.

Содержание 3500

Страница 1: ...Instructions 2 Before Assembly Assembling the Scale Attach Control Panel Box Leveling Platform Battery Charging 3 Charging the Battery Product Description 3 Control Panel Box Special Features Specifications Operating Instructions 4 Using the Zero Function Taking a Weight Measurement Care and Maintenance 5 Cleaning Calibration Troubleshooting Guide 5 Parts Lists 6 Warranty 7 OPERATOR S MANUAL ...

Страница 2: ...ientation of new personnel Use only the appropriate connecting cable designed specifically for use with this monitor If the cabling provided with this unit becomes cracked or damaged it should be repaired or replaced immediately If the scale becomes damaged it should not be operated until properly serviced All repairs should only be performed by authorized Tanita Corporation service personnel Scal...

Страница 3: ...read the control cable with connector through the right handlebar mounting hole in the wheel cover Insert the cable into the keyway in the mounting hole Fig 1 4 Lay the platform down Insert the handlebar into the 2 two mounting holes in the wheel cover Make sure the control cable is seated in the mounting hole keyway Fig 2 5 Line up the holes in the handlebar with the holes in the wheel cover Secu...

Страница 4: ...o be recharged Charging the Battery A Plug the battery charger power cord into the 3 prong jack on control panel box Secure by turning the knurled ring B Plug the battery charger into a 120 VAC outlet C The battery will fully charge in 8 hours THE SCALE MAY BE USED WHILE THE BATTERY IS CHARGING NOTE The specially designed charger built into this scale prevents the battery from being overcharged Us...

Страница 5: ...12 5 lb 5 6 kg OPERATING INSTRUCTIONS Using the Zero Function When using the Stand on Scale the weight of any object accompanying the patient blanket etc must be taken into account during a weighing Weigh the object s by themselves to determine their weight and subtract this weight from the combined weight of patient and object s METHOD 1 1 Scale is off 2 Place the object blanket etc on the scale ...

Страница 6: ...eatable reading In this case a series of moving dashes continues to appear on the console rather than a weight reading If moving dashes persist activate the Average Lock by depressing the appropriate Pounds or Kilos switch to obtain an average reading of the patient s weight Note This reading is now an instantaneous average of weights that the computer calculates 5 After weight is obtained and rec...

Страница 7: ...1 3 CAP ELE 470UF 25V C1 C7 C12 1 CERAMIC RES 4 00MHz X1 4 TRANS PNP PWR 2N6728 Q1 Q2 Q3 Q4 2 TRANS PNP PWR TIP32 Q5 Q7 1 TRANS PNP GEN MPS3702 Q8 13 TRANS NPNGEN MPS3704 Q6 Q9 Q10 Q11 Q12 Q13 Q14 Q15 Q16 Q17 Q18 Q19 Q2 4 DIODE SIGNAL 1N4148 CR1 CR7 CR14 CR15 1 DIODE ZENER 9 1V 1N4739A CR2 10 DIODE REC 600V 1N4005 CR3 CR4 CR5 CR6 CR8 CR9 CR10 CR11 CR12 CR13 2 IC REG 5V 7805 U1 U2 1 IC MICROPROCESS...

Страница 8: ... by the breach of an express or implied warranty or condition Except to the extent prohibited by applicable law any implied warranty of merchantability or fitness for a particular purpose is limited in duration to the duration of the above warranty Tanita disclaims all other warranties or conditions express or implied statutory or otherwise Some provinces or jurisdictions do not allow the exclusio...

Отзывы: