1.8
Installation
The PCU1 and associated loading units are designed for use in an indoor environment.
The unit must be connected to an installation category II supply (wall socket). Before
installation, a suitable mains connector must be connected to the 3-core mains input cable
provided. The connector and supply must be rated for 32A for output currents up to 3000A
and 50A for output currents up to 5000A. It is also essential to ensure that the unit is
adequately earthed.
WARNING
Ensure that the unit is adequately earthed. If necessary, an extra earth
connection may be made between the earth terminal on the front panel of the
control unit and a suitable low impedance local earth.
WARNING
Ensure that the unit is connected to an appropriate supply by a suitably rated
connector.
WARNING
The PCU1 series control units and NLU series loading units are heavy. Avoid
lifting where possible. Ensure that two people are used to lift the units and that
appropriate manual handling techniques are used.