8

1.1

Electrical Specification

1.1.1

The supply voltage will be as follows:

220/240V 1 phase 50/60Hz+10%
115V 1 phase 50/60Hz +10% -6%

1.1.2

The minimum supply requirements are 1200VA:

1.1.3

The output voltage, current and duty cycle are as follows:

Output Voltage

Output Current 1

Output    Current 2

KV30-40

30kV

20mA    40mA

KV15-80

15kV

40mA    80mA

Output current 1

is the continuous current rating.

Output current 2

is based on a duty cycle of 5 minutes ON load followed
by 15 minutes OFF load.

1.1.4

One side of the high voltage transformer output is connected to the earth system of
the equipment.

1.1.5

System Interlock:

The equipment is fitted with a zero volt interlock system on the controlling regulator.
This interlock prevents the output being energised unless the controlling regulator is in
the minimum output position.

 

1.1.6

An earthing hook is provided which is stored in the lid of the test set.

Содержание KV15-18

Страница 1: ...h Voltage AC Test Set Type KV30 40 KV15 18 DESIGNED AND MANUFACTURED BY T R Test Equipment Limited Keens Lane Worplesdon Guildford Surrey GU3 3LS United Kingdom Telephone 01483 235757 e mail sales trt...

Страница 2: ...2...

Страница 3: ...he equipment that may be connected to a local earth Ensure that the product is properly grounded before making any connections to inputs or outputs Terminal ratings must be observed to prevent fire ha...

Страница 4: ...4...

Страница 5: ...5 CONTENTS 1 Description of Equipment 2 Operating Procedures 3 Maintenance Instructions 4 Standard Accessories 5 Overall Performance Specification 6 Calibration Results 7 Test Certificate 8 Revision...

Страница 6: ...6...

Страница 7: ...7 1 DESCRIPTION OF EQUIPMENT 1 1 Electrical Specification 1 2 Output Voltage Control 1 3 Overload Protection 1 4 Metering 1 5 Construction...

Страница 8: ...rrent 1 is the continuous current rating Output current 2 is based on a duty cycle of 5 minutes ON load followed by 15 minutes OFF load 1 1 4 One side of the high voltage transformer output is connect...

Страница 9: ...is controlled from zero by means of a regulating transformer The transformer is operated by means of a control knob which is located on the front panel of the control unit To increase the output volta...

Страница 10: ...3 3 Variable overload The variable overload protection system senses current changes in the high voltage circuit The trip level is adjusted by means of a selector switch on the front panel of the cont...

Страница 11: ...is provided with the following metering a A moving coil dual range kilovoltmeter monitoring the output voltage from the equipment b A moving coil single range milliammeter monitoring the secondary cur...

Страница 12: ...12 1 5 Construction 1 5 1 The equipment is housed in a robust case with recessed carrying handles All the controls and metering are located on the front panel together with all external connections...

Страница 13: ...13 2 OPERATION 2 1 Connections 2 2 Front Panel Functions 2 3 Method of Operation...

Страница 14: ...earth to the earth terminal on the equipment 2 1 3 Connect the earth hook lead to the earth terminal on the test set and place the hook on the test object s high voltage terminal 2 1 4 IMPORTANT The o...

Страница 15: ...earth return test object earth stick IMPORTANT There must be 20cm minimum clearance between the unscreened section of HV output cable and earth Note The test object is shown earthed via the earth stic...

Страница 16: ...t Switches push button Green OFF HV output when pressed d HV ON illuminated Indicates HV output is ON when lit Switches push button Red ON HV output when pressed e Trip level mA switch Selects output...

Страница 17: ...el Layout EARTH RATING MAINS FUSE IN VOLTAGE SELECTOR DANGER HIGH VOLTAGE Ma kV TRIP X 0 5 X 1 MILLIAMPS RESET 10 KV RANGE ON HV OUTPUT OFF 0 100 POWER OUTPUT CONTROL TYPE KV30 40 T R TEST EQUIPMENT g...

Страница 18: ...push button will go out thus indicating that the circuit is armed 2 3 5 Select the desired voltage range on the kV meter and set the desired overload mA trip level 2 3 6 Before commencing testing ensu...

Страница 19: ...commended If however the track is badly corroded it should be cleaned with very fine sand paper ensuring a flat surface is restored Remove particles with a fine brush and finally cleanse with an alcoh...

Страница 20: ...5A 1 inch b 1 off F10A 1 inch 4 2 The following items are provided with the equipment a Mains input lead b 2 keys for mains ON OFF switch c 5 metre earth lead fitted with clip and tag d Type ES30 40 e...

Страница 21: ...frame 5 3 Accuracy of instruments KV meter 1 5 at full scale mA meter 1 5 at full scale 5 4 H V Transformer 2 5kV RMS for 1 minute between primary and earth 2 5kV RMS for 1 minute between earthy end o...

Страница 22: ...22...

Страница 23: ...er Equipment s mA meter Standard mA meter 5 0 20 High Range x1 10 0 40 15 0 60 2 5 80 Low Range x0 5 5 00 100 7 5 Unit Type KV30 40 Range Equipment s KV meter Standard KV meter Equipment s mA meter St...

Страница 24: ...24...

Страница 25: ...PE KV15 80 PRODUCT SERIAL NUMBER TEST PROCEDURE As routine test sheet KV15 80 rts doc TESTED BY PASSED BY DATE PRODUCT TYPE KV30 40 PRODUCT SERIAL NUMBER TEST PROCEDURE As routine test sheet KV30 40 r...

Страница 26: ...26...

Страница 27: ...VISION Product Type High Voltage AC Test Set KV30 40 KV15 18 File KV30 40 and KV15 80 manual v10 doc DJ1 Author D Buckle Issue Date 10 12 06 2001 Modified By D Buckle Checked By I D W Lake Date 12 06...

Страница 28: ...28 Drawings Required A2 000308 latest issue A2 000542 latest issue...

Отзывы: