Specifications
Model
RTR501B / 501BL
RTR502B / 502BL
RTR503B / 503BL
RTR507B / 507BL
Measurement
Channels
Temperature 1ch
Temperature 1ch
Temperature 1ch, Humidity 1ch
Temperature 1ch, Humidity 1ch
Sensor
Thermistor (Internal)
Thermistor
Thermistor
Polymer Resistance
Thermistor
Polymer Resistance
Measurement
Range
-40 to 80℃
-60 to 155℃
0 to 55℃
10 to 95%RH
-25 to 70℃
0 to 99 %RH (*1)
Accuracy
Avg. ±0.5℃
Avg. ± 0.3℃
at -20 to 80℃
Avg. ±0.5℃
at -40 to -20℃, 80 to 110℃
Avg. ± 1.0℃
at -60 to -40℃, 110 to 155℃
Avg. ± 0.3℃
±5 %RH
at 25℃, to 0%RH
±0.3°C
at 10 to 40℃
±0.5°C
Others
±2.5 %RH
at 15 to 35℃,
30 to 80%RH
Measurement
Resolution
0.1℃
0.1℃
0.1℃
1 %RH
0.1℃
0.1 %RH
Responsiveness
Thermal Time Constant:
Approx. 15 min.
Approx. 25 min. (L Type)
Response Time (90%):
Approx. 35 min.
Approx. 47 min. (L Type)
Thermal Time Constant:
Approx. 30 sec. (in air)
Approx. 4 sec. (in agitated water)
Response Time (90%):
Approx. 80 sec. (in air)
Approx. 7 sec. (in agitated water)
Response Time (90%): Approx. 7 min.
Response Time (90%): Approx. 7 min.
Logging
Capacity
16,000 readings
16,000 readings
8,000 data sets (One data set con-
sists of readings for all channels in
that type of unit.)
8,000 data sets (One data set consists of
readings for all channels in that type of unit.)
Recording
Interval
1, 2, 5, 10, 15, 20, 30 sec. 1, 2, 5, 10, 15, 20, 30, 60 min. (Total of 15 choices)
Recording
Mode (*2)
Endless: Overwrite the oldest data when capacity is full
One Time: Stop recording when capacity is full
Communica-
tion Interfaces
Short Range Wireless Communication
Frequency Range: 902 to 928 MHz
RF Power: 7 mW
Transmission Range: About 150 m if direct and unobstructed.
Bluetooth 4.2 (Bluetooth Low Energy) (*3)
Optical Communication
Power
Lithium Battery LS14250 x 1
L Type: Large Capacity Battery Adaptor Kit (RTR-500B1) x 1 (*4)
External Power Adaptor Kit RTR-500A2 (*5)
Battery Life (*6)
About 10 months L Type: About 4 years
Dimensions
H 62 mm x W 47mm x D 19 mm L Type: H 62 mm x W 47 mm x D 46.5 mm (excluding protrusions and sensor)
Weight
Approx. 50 g, L Type: Approx. 65 g
Operating
Environment
-40 to 80℃
Wireless Communication -30 to 80℃
Waterproof
Capacity
IP67 (immersion proof)
IP64 (splash proof / rated for use in daily life) (*7)
Compatible
Base Units
RTR500BC, RTR500BW
RTR-500DC, RTR-500MBS-A, RTR-500NW/AW (*8)(*9)
RTR-500 (*9)
*1:
When continually used in environments with temperatures above 60 ̊C, accuracy of humidity measurements will decrease over time. Also, humidity cannot be measured at temperatures below -20 ̊C.
*2: Only “Endless” is available when using the RTR500BW, RTR-500NW/AW or RTR-500MBS-A as a Base Unit.
*3: Bluetooth is available when using the RTR500BW as a Base Unit and making device settings in the mobile app (T&D 500B Utility).
*4: When using RTR-500B1 it is necessary to purchase Lithium Battery (LS26500). For details, contact your local authorized distributor.
*5: RTR-500A2 should not be used with the RTR501B, as it will cause the RTR-501 to display a higher than actual temperature reading of up to 3°C.
*6: Battery life varies depending upon the ambient temperature, radio environment, frequency of communication, recording interval, and quality of the battery being used. All estimates are
based on operations carried out with a new battery and are in no way a guarantee of actual battery life.
*7: This is the waterproof capacity of the data logger with the sensor connected. Note that the temperature-humidity sensor is not water resistant.
*8:
A firmware update is required.
*9: A software update is required.
The specifications listed above are subject to change without notice.
How to Read the LCD Display
When being used in very hot or cold environments the display may become difficult to read. This is not a malfunction.
[REC] Mark
Shows the recording status.
ON: Recording in progress
BLINKING: Waiting for programmed start
OFF: Recording stopped
[ONETIME] Mark Shows the recording mode.
ON: One Time
OFF: Endless
Battery Warning
Mark
Indicates it is time for the battery to be replaced.
Measurement /
Message Area
Measurements or operational messages are shown
here.
When Recording Mode has been set to “One Time” and
the logger reaches its logging capacity, recording will
automatically stop and in the LCD the measurement
and the word [FULL] will alternately appear.
The measurement and the word [SEnd] will alternately
appear when data is being sent via wireless communi-
cation to the Base Unit. Recording will continue during
wireless transmission.
[Ir] Mark
Indicates that the logger (Remote Unit) has not been
registered to the Base Unit or the wireless communi-
cation is stopped (inactive radio).
[COM] Mark
Indicates that the device is in Bluetooth communication.
Unit of Measure-
ment
Shows the unit of measurement.