Reproducibility
The reproducibility of extracted lysates on the same machine (run-to-run precision) is
the variability of the results from repeated measurements. The coefficient of variation
of nucleic acid extraction concentration is less than 5%. It is in the range of acceptable
statistical analysis.
▓
Nucleic Acid Yield
N=8
Cycle 1
Cycle 2
Cycle 3
Cycle 4
Average
Yield
N=128
Standard
Deviation
(SD)
Coefficient
of
Variation
(CV%)
Average
Yield
(ng/
l)
25.3
24.7
25.3
26.5
25.0
0.588
2.35
Standard Deviation
(SD)
0.005
0.005
0.007
0.006
Coefficient
of
Variation
(CV%)
2.80
1.83
1.79
2.77
▓
Ratio (A260/A280)
N=8
Cycle 1
Cycle 2
Cycle 3
Cycle 4
Average
Ratio
N=128
Standard
Deviation
(SD)
Coefficient
of
Variation
(CV%)
Average Ratio
1.87
1.86
1.89
1.89
1.87
0.014
0.79
Standard Deviation
(SD)
0.005
0.005
0.007
0.006
Coefficient
of
variation
(CV%)
0.27
0.24
0.37
0.34
Содержание Maelstrom 4800 LH
Страница 1: ...TANBead Nucleic Acid Extractor Non Sterile Maelstrom 4800 LH User Manual...
Страница 7: ...6 Equipment list No Item 1 Power cord 2 A plate of spin tips 3 Magnetic separator 4 User Manual...
Страница 9: ...8 Side view of extraction platform Shield plate...
Страница 10: ...9 Regulators of the sensors for detecting falling spin tip Control panel Buttons on control panel...
Страница 12: ...11 Set the Temperature UV LAMP Confirm the Selected Program Mount the Spin Tips...
Страница 13: ...12 Eject the Spin Tips STATUS RUNNING STATUS PAUSE STATUS PAUSE STATUS END...
Страница 17: ...16 NOTE 1 NOTE 2 Definition of plate number...
Страница 22: ...21 6 2 Communication between M4800 LH and Liquid Handling System...
Страница 23: ...22 6 4 Communication CRC Cyclical Redundancy Check...