4. Data Storage Function (use this function with optional Built-in Clock Board)
APM-7 has a data storage function. Up to 50 successive data can be memorized. (
*5
)
After APM-7 detects a flash point, the data will be stored automatically.
If Built-in Clock Board is not installed, actual date will not be memorized in
DATE
column.
The recalling procedure is as follows.
Press the
MODE
switch and
RESET
switch, then consecutively release the
RESET
switch and
then the
MODE
switch. The following screen will appear:
Fig. 10 Data Rree
Fig. 10 Data Recall Screen
By pressing Up-Down key, the screen will be scrolled.
If needed, press
FUNCTION
key to print out all data. (when optional printer is connected)
*5
: Following data will not be stored.
-
used special mode, simulated TAG mode and user custom mode
-
flash is detected at first application
-
flash is detected by manual application
-
flash is detected at a temperature which is greater than 28
o
C (50
o
F) in Procedure A & B /
30
o
C in Procedure C, above the temperature of the first application of the ignition source,
or when a flash point is detected at a temperature witch is less than 18
o
C (32
o
F) in
Procedure A & B / 16
o
C in Procedure C, above the temperature of the first application of
the ignition source.
(New Data)
(Old Data)
APM DATA SAMPLE ID FLASH.P
01 03/14 09 GAS OIL 60.5
o
C
02 03/13 18 GAS OIL 60.0
o
C
03 03/13 17 N-CETANE 134.0
o
C
Month
Day
Hour
Sample ID
Corrected
Flash Point
Number
01: latest
50: oldest