OPERATING THE SYSTEM 

 
When the bath is ready for use it can be switched on by pressing the mains switch.  
 
The electronics are suited for both 50 and 60 Hz. Check whether the voltage of the system conforms to the 
mains voltage. 
 

6.1 

Quick Start 

 
To  start  operating  the  bath  in  a  quick  way  do  the 
following: 
 
Fill  the  bath  with  fluid  as  indicated  in  "Bath  fluid  level", 
page 7, 
Place the power plug, connect to mains socket, 
Switch the bath on using the mains switch, 
Select appropriate set point, 
 
PID settings 
All measuring results have been acquired using following 
PID settings: 
 
Pb 

:= 

25 

:= 

16 

:= 

 
Under  different  settings  its  possible  to  achieve  even 
better values by trimming the PID settings.  
 
Place  the  bath spirit  level.  The  four  supporting  feet  can 
be turned in and outwards for exact adjustment. 
 

6.2 

Menu Items 

 
Use the turn-push button to select a menu item and select 
the item by pressing the button. After pressing a     sign 
appears next to the value indicating the value can now be 
changed.  Pressing  the  button  again  activates  the  value 
immediately. When the value is altered but the button is 
not pressed the value will be accepted and stored after 5 
seconds automatically. The menu returns back to normal 
operating mode. 
 
 

Menu item "Set Point" 

 
Select the set point temperature. Resolution is 0.01°C +/- 
the system accuracy. 
 
 

Menu item "Offset" 

 
The  temperature  displayed  can  be  increased  or 
decreased  with  an  offset  ranging  from  +5.00  down  to  -
5.00 °C in steps of 0.01°C. This way the temperature  

Spirit level 

Содержание TV12LT

Страница 1: ...3 tv12ltman Rev 3 01 UK 0821 E mail info tamson com Website www tamson com Tamson Instruments bv Page 1 30 ISO 9001 2015 NL PRO 238239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350 370...

Страница 2: ...12 4 2 CONSTRUCTION 12 4 3 STIRRER 12 4 4 TEMPERATURE CONTROL AND SETTING 12 4 5 FRONT PANEL 13 4 6 OVERVIEW MENU ITEMS 13 5 SAFETY SYSTEMS 15 5 1 MECHANICAL OVER TEMPERATURE PROTECTION 15 5 2 LEVEL P...

Страница 3: ...e to the equipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect the apparatus at the main s socket Proper sk...

Страница 4: ...the equipment or its accessories If a malfunction occurs consult section TROUBLE SHOOTING page 24 at the end of this manual If problem persists call your supplier or Tamson Instruments bv Never opera...

Страница 5: ...companies generally will not honor a claim for damage if the respective packing material is not available for examination The shipment contains at least the bath as mentioned in the delivery checklist...

Страница 6: ...well above the maximum operating temperature Fluid must not be aggressive when in contact with stainless steel brass glass and silicon sealing Extreme HAZARD Do not use methanol at temperatures above...

Страница 7: ...e bath liquid ignites under these conditions The heating element must be fully covered by the bath fluid The heating element will be damaged when not fully submerged in the bath fluid 3 1 Drain bath f...

Страница 8: ...5 RABO 0160100046 Chamber of commerce 27 16 95 41 3 2 Bath fluid level Minimum level The heating element must be fully covered by the bath fluid The heating element will be damaged when not fully subm...

Страница 9: ...t fluid from the overflow outlet entering the backside of the TV12LT For this reason the overflow outlet must be connected to a waste container When the bath is working at high temperatures tubing and...

Страница 10: ...in two functions When cooling down to the set point temperature maximum cooling capacity is used Whe the set point temperature is reached the bath will start to tune for stable control It also switch...

Страница 11: ...1 2015 NL PRO 238239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 3 5 Connecting Before plugging TV12LT into the mains socket...

Страница 12: ...emperature control and setting The bath temperature is regulated using a Pt 100 temperature probe which is connected to a microprocessor module The advanced electronic control system continually compu...

Страница 13: ...hermostat 8 Turn push button Use well grounded mains Before plugging the TV12LT into the mains socket make sure the voltage of the bath corresponds to the local voltage The front panel layout shows th...

Страница 14: ...or is restarted the displayed value increases to a stable readout appears after a few seconds Ad 2 The controller calculates every second the amount of power which should be applied for stable contro...

Страница 15: ...h temperature drops approximately 10 C below the pre set temperature To continue normal operation the bath has to be switched off and on again 5 1 Mechanical over temperature protection Turn the therm...

Страница 16: ...t point PID settings All measuring results have been acquired using following PID settings Pb 25 I 16 D 0 Under different settings its possible to achieve even better values by trimming the PID settin...

Страница 17: ...ath This menu item enables or disables the heater Standard value On Menu item Time const Used to select time to tune The option precise has to be used to reach maximum temperature accuracy Options are...

Страница 18: ...O 0160100046 Chamber of commerce 27 16 95 41 Pb proportional band Proportional value P is found to be 100 Pb Off Standard value On Menu item Temp units C F Standard value C Menu item Baudrate 300 600...

Страница 19: ...350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 PID settings Following shows influence on stability with different PID settings Temperature 75 C Proportional band Pb 25 Integrator 16 Dif...

Страница 20: ...rmined with the use of the Ziegler Nichols method described below With the process at its normal process value Set the integral time Ti and the derivative Time Td to off Check if the Lcb and Hcb are s...

Страница 21: ...Chamber of commerce 27 16 95 41 8 Maintenance Very important Keep the condenser free from dus Remove side panel Use vacuum cleaner to remove dust Dust builds up depending on environment Regularly che...

Страница 22: ...r is not turning The motor fuse is activated Restart the motor by pressing the motor fuse This fuse is located at the backside of the top casing Also check viscosity of the bath fluid High viscosity w...

Страница 23: ...he bath fluid used is too high The maximum viscosity of the bath fluid lies below 10 mm s at the operating temperature but is preferably less than 3 mm s If the viscosity of the bath fluid used is too...

Страница 24: ...elements are submersed 9 1 Over temperature safety thermostat When the thermostat is activated the entire bath is switched off with the exception of the controller the red system error LED will light...

Страница 25: ...mperature 5 C Proportional band Pb 25 Integrator 16 Differentiator 0 Min max 0 013 C Temperature 20 C Proportional band Pb 25 Integrator 16 Differentiator 0 Min max 0 015 C 10 2 Cooldown 10 3 Settling...

Страница 26: ...mostatic baths contain heating Simple bath constructions have a single heater and stirrer for circulation This causes random energy distribution and poor homogeneity The heating energy is distributed...

Страница 27: ...O 0160100046 Chamber of commerce 27 16 95 41 When aligning viscometers or sensors in horizontal position they will all have the same temperature Front view Side vie 1 2 T 3 Side view 4 T Devices have...

Страница 28: ...Amp 25T1123 n a Compressor 230V 50Hz 25T1127 n a Compressor 230V 60Hz 25T1126 Compressor 115V 60Hz 25T0351 DC Heater W 25T0404 Window heating 4x 115V placed in series 24T8581 Over temperature protecti...

Страница 29: ...tive of the European Parliament and of the Council of 17 May 2006 42 EC 2nd Edition June 2010 p d EN 60204 Machinery Directive and Safety requirements p pi e EN60950 1 Low Voltage Directive p f EN6100...

Страница 30: ...uarantee Conditions of use and suitability of the product for particular uses are beyond our control all risks of use of the product are therefore assumed by the user We expressly disclaim all warrant...

Отзывы: