tv2000-4000mkII-4000DC-man.doc Rev. 2.3 UK 9-15
Never empty hot oil into a plastic container.
4.3.1
Maximum level
If the fluid level is too high, it will leave the bath via the
overflow outlet (14mm or 3/8" outher diameter pipe, no
thread). Prevent fluid from the overflow outlet entering
the side of the bath. For this reason the overflow outlet
must be connected to a waste container.
When the bath is working at high temperatures, tubing
and waste container must be chemical resistant and
able to withstand high temperatures (>230°C / 446°F).
4.3.2
Drain bath liquid
The bath can be emptied via the drain tap located at the
backside of the apparatus. For safety reasons the tap
can only be opened by using a screwdriver. The thread
inside the tap is 3/8".
Take necessary precautions against fire hazard
when removing flamable bath fluid.
When removing bath fluid do not inhale toxic vapor.
Always use appropriate ventilation.
Handle old bath fluid as toxic waste.
Cool down bath fluid to ambient before removing.
Connect outlet
to waste
container
PREVENT OVERFLOW
Do not drain
hot liquid !!
Oil
PLASTIC
CONTAINER
USE
SCREWDRIVER
TO OPEN DRAIN
3/8" BSP inside
thread
Tubing and container:
- Chemical resistant
- Withstand high temperature
Oil