background image

 

 
 
 
 

 

Page 1/30 

 

Van ’t Hoffstraat 12 
2665 JL Bleiswijk, The Netherlands 
T. 31 (0) 10 522 43 73 
F. 31 (0) 10 521 19 41

 

 
 
E-mail: [email protected] 
Website: www.tamson.com 

 

VAT: NL 80 66 34 984 B01 

Bank account no.: Rabobank 16 01 00 046 

Chamber of commerce 27 16 95 41 

IBAN Code: NL95 RABO 0160100046

 

 

Tamson Instruments bv 

tv2000-4000mkII-4000DC-man.doc Rev. 2.3 UK 9-15 

 

 
 
 
 
 

Manual 

TV2000-MKII 
TV4000-MKII 

TV4000-MKII DC 

Содержание PMT TV2000MKII

Страница 1: ...21 19 41 E mail sales tamson com Website www tamson com VAT NL 80 66 34 984 B01 Bank account no Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code NL95 RABO 0160100046 Tamson Instruments...

Страница 2: ...ITEMS 14 6 2 SAFETY THERMOSTAT 15 6 3 DISPLAY 15 6 4 QUICK START 16 7 MENU ITEMS 17 7 1 1 Menu item Set point 17 7 1 2 Menu item Offset 17 7 1 3 Menu item Max Power 17 7 1 4 Menu item Boost heater 17...

Страница 3: ...VAT NL 80 66 34 984 B01 Bank account no Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code NL95 RABO 0160100046 Tamson Instruments bv tv2000 4000mkII 4000DC man doc Rev 2 3 UK 9 15 9 ADD...

Страница 4: ...damage to the equipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect at the mains socket of the apparatus P...

Страница 5: ...sonnel Avoid damage to the equipment or its accessories through incorrect operation 4 INSTALLATION 4 1 Important Tamson Instruments b v is not responsible for any consequential damage or harm caused b...

Страница 6: ...d as long as the viscosity of the fluid is low enough at the operating temperature Following products can not be used as bath fluid Distilled water Aggressive fluids when in contact with stainless ste...

Страница 7: ...oil 50 250 transparent 150cSt 50 C 122 F 6cSt 150 C 302 F F P 315 C 599 F Lifetime at 200 C 392 F is 1jr at 250 C 482 F 1200hrs when not polluted The oil used has a limited lifetime The type brand an...

Страница 8: ...c Rev 2 3 UK 9 15 Using the bath for long periods at low temperatures will cause condensation on the glass window When operating at high temperatures the lid of the bath top plate and the window secti...

Страница 9: ...UK 9 15 Use proper ventilation Heated bath fluids can cause toxic fumes Remove fumes from hot oil or bath medium Use fume cabinet or proper air ventilation When changing the bath fluid from water to o...

Страница 10: ...to a waste container When the bath is working at high temperatures tubing and waste container must be chemical resistant and able to withstand high temperatures 230 C 446 F 4 3 2 Drain bath liquid Th...

Страница 11: ...at lower temperatures it will be necessary to pass cooling liquid from a cooling circulator TLC10 or TLC15 or tap water through the cooling coil As an example to operate at 40 C 104 F with an ambient...

Страница 12: ...temperature 5 1 Construction The TAMSON baths are constructed entirely from corrosion resistant materials such as stainless steel and PTFE Teflon A thick layer of rock wool between the inner bath and...

Страница 13: ...nd the clock working of the bath These features protect the equipment the bath fluid temperature and the workplace A mechanical over temperature protection thermostat 1 will automatically switch off t...

Страница 14: ...6 1 Overview menu items Set point Offset press 5 00 5 00 C resolution 0 01 C Max Power press low med hi max Boost heater press on off Time const press fast medium slow precise PID parameter PID set 1...

Страница 15: ...d on again When switching off wait 5 seconds before switching on again 6 3 Display 1 Temperature readout 2 Applied percentage of power 3 Operating mode 4 Indicator alarm high alarm low control stable...

Страница 16: ...start operating the bath in a quick way do the following Fill the bath with liquid as indicated place the power plug connect to mains socket switch on the bath using the mains switch select appropria...

Страница 17: ...th temperature It can be set in steps of 0 01 C 7 1 2 Menu item Offset The temperature displayed can be increased or decreased with an offset ranging from 5 00 down to 5 00 in steps of 0 01 C The offs...

Страница 18: ...gh alarm Optional min SP to max SP resolution 0 1 C No hardware connected display function only Optional 7 1 9 Menu item PID parameter PID set 1 First set of parameters PID set 2 Second set of paramet...

Страница 19: ...et of 0 005 C The value of 0 005 C is added to the selected SP value The SP can be selected with 0 01 C accuracy 7 1 14 Menu item Restart Restarts system and activates tuning 7 2 PID settings Default...

Страница 20: ...NL 80 66 34 984 B01 Bank account no Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code NL95 RABO 0160100046 Tamson Instruments bv tv2000 4000mkII 4000DC man doc Rev 2 3 UK 9 15 7 3 TV400...

Страница 21: ...of the Ziegler Nichols method described below The PID parameters can also be determined with the use of the Ziegler Nichols method described below With the process at its normal process value Set the...

Страница 22: ...the temperature of the bath has been lowered with approximately 20 degrees the thermostat re sets itself automatically however to continue normal operation the bath has to be switched off and on again...

Страница 23: ...irculating system is incapable to mix the bath fluid thoroughly which may result in poor stability Power is limited to a very low level Position of PT100 has been changed The tip of the PT100 must be...

Страница 24: ...hen PV is in the region of 100 C PV 0 C StableLimitHigh is defined as 0 05 When the average PV deviation calculated from a sliding window of 255 seconds is smaller than StableLimitHigh the ST command...

Страница 25: ...decimal value with 7 decimal digits e g 12 345342C works for degrees Centigrade and degrees Fahrenheid Only first three are significant When reading is in degree Fahrenheit character F indicates the...

Страница 26: ...lete TV2000MKII 25T0210 25T0230 Heater 1500W 25T0210 25T0210 Boost Heater 25T1344 Capacitor 5 5uF 24T3312 Capacitor 10uF 24T8081 Motorfuse 0 3 Amp 24T8080 Motorfuse 0 6 Amp 24T8541 Mains switch 25T124...

Страница 27: ...an doc Rev 2 3 UK 9 15 9 ADDENDUM TECHNICAL REFERENCE 9 1 Technical Drawing TV2000 TV4000MKII Motor fuse Heater Boost Over temperature Thermostat Fan PT100 Neutral RS232 female Flatcable powerboard He...

Страница 28: ...c Rev 2 3 UK 9 15 9 2 Technical drawing TV4000MKII DC Motor fuse Heater Boost Over temperature Thermostat Fan PT100 Neutral RS232 female Flatcable powerboard Level detect 12V DC PE Phase PE Mains Indi...

Страница 29: ...e 2006 95 EC relative to low voltage The products are in conformity with the following specifications EMC 2004 108 EG Conducted emission EN55016 2 1 EN61326 A1 Radiated emission EN55016 2 3 EN61326 A1...

Страница 30: ...every kind and nature including warranties of merchantability and fitness for a particular purpose in respect to the use or suitability of the product Nothing is intended as a recommendation for uses...

Отзывы: