17
Installation
■ Installing the Main Unit
● Use this instrument under the following environmental conditions:
• Temperature:
+10 to +35 degrees celsius
• Humidity:
30 to 75%
● Install the instrument correctly by following the procedure below.
1.Remove the instrument from the packing box.
・Do not use a box cutter or other sharp tools to
open the box.
・Check the contents of the box.
● When fully assembled, the OM-8 weights more than 80kg. Please assemble it at or near the site whereby it will be
used.
● When assembling it in a nearby place, check in advance that there are no highly uneven areas in the floor between
there and the site of use.
● At least two people are required for assembly. One person should not try to do it alone.
● When installing the coupling unit, do not turn the balance adjustment handle, as that could upset the adjustment
of the counter-balanced arm and adversely affect the functioning of the unit.
● When installing the microscope unit, do not turn the balance adjustment handle, as that could upset the
adjustment of the counter-balanced arm and adversely affect the functioning of the unit.
● When moving the OM-8, fold the arms as shown below, and then tighten all locking knobs and secure all the
parts.
● Do not move the OM-8 while the arms are extended. It can cause its overturning.
● Do not move the OM-8 while the casters are locked. It can cause its overturning.
● Strong jolts can affect the functioning of the OM-8 and cause other problems, so be sure to be careful when
moving it.
● Assembling the Stand Unit
1.Immobilize the base by lowering the two
stoppers on the base's two stopper-attached
casters.
OM8-T26
Lower to Lock
CAUTION
OM8-T25̲2
OP E R A T I NG M I C R OS C OP E
OP E R A T I NG M I C R OS C OP E
W A R NI NG L OC K
○
×