14

1. Open the OSD menu

To open the TAGARNO On Screen Display Menu, press and hold the 
center button for approximately 5 seconds.

2. Choose lens 

Set the current close-up lens by using the left/right arrows on the 
CONTROL BOX XKEY. On a TAGARNO FULL HD system, the choice 
naturally is between LENS +2, +3, +4, +5 or +10. LENS +4 is factory 
default. Continue using the down arrow.

3. Choose monitor width

Press the center button to set up the width of the monitor. You need 
to measure the horizontal width of the monitor panel on your monitor, 
Select one digit at a time by using the left/right arrows and adjust the 
digits with the up/down arrows. Switch between millimeters and inches 
via the up/down buttons. Monitor 24” FHD is factory default. To store 
change in settings press the center button.

4. Set start magnification

Select which magnification level you need your system to use as start 
up level. Press the center button and select one digit at a time in the 
bottom of the page, by using the left/right arrows and adjust the digits 
with the up/down arrows. By pressing the center button one more time, 
you have selected the values chosen. Press the down arrow to choose 
format.

When turning on your FULL HD system, you are informed of the 
given units, serial number (SN), version and resolution.

ON SCREEN DISPLAY (OSD)

5. Choose format 

You have the option to switch between different video formats, 1080p 
60fps being the highest quality. Select the required format by pressing 
the center button, and use arrow keys left/right to select between 6 
different formats. 

6. Select OSD presets

You need to choose between respectively ALWAYS ON, OFF or 
TIMEOUT using the left/right arrows in order to have the OSD Menu 
displayed continuously, never or for 3 seconds at the time. 

 

Continue by pressing the down arrow.

7. Color boost

The OSD menu gives you the possibility to choose between two color 
settings; Color boost on or off, by using the left/right arrows. Which 
setting chosen, is a matter of individual preferences and the object 
projected on screen. Continue by pressing the down arrow.

8. Flip picture

In the menu you have the possibility to rotate the screen image 180 
degrees, or choose the standard view by using the left/right arrows. 
Choose between the two views by pressing the left/right buttons. 
Continue by pressing the down arrow.

9. White balance

Choose between these settings: AUTO and POWER UP by using the 
left/right arrows. In AUTO mode the white balance is continuously 
adjusted to achieve the best color reproduction. In POWER UP mode 
the white balance calibration is performed only once when the system 
is turned on. In this mode it’s important that a white sheet of paper 
is visible in the field of view when the microscope is switched on. 
Continue by pressing the down arrow.

10. Save presets

To save your presets and exit the menu, press the center button and 
thereby return to the image displayed on the monitor, using your 
recently saved presets. 

If you have changed the various formats, your FULL HD system needs 
to be restarted. This will be indicated in the bottom if needed.

OPERATION (3/3)

Содержание ZAP STAND

Страница 1: ...D UPGRADE KIT 4 6 ASSEMBLING ZAP WITH Ø76 ADAPTER KIT 7 7 CONNECTING 10 8 OPERATION 12 9 TECHNICAL SPECIFICATIONS 15 10 PC REQUIREMENTS 17 11 MAINTENANCE 18 12 WARRANTY 18 13 ERGONOMICAL RECOMMENDATIONS 18 14 DECLARATION OF CONFORMITY 19 15 COMPLIANCE STATEMENTS 20 MANUAL TAGARNO ZAP STAND Go to tagarno com productmanuals to find manuals and tutorial videos ...

Страница 2: ...mechanical parts and a power supply The product is intended for marketing worldwide and is designed for manual visual inspection INTENDED USE Read all safety information before you use the product Please pay attention when you see a warning label on the product This product is for indoor use only You must not discard this electrical electronic product in domestic household waste Please dispose at ...

Страница 3: ...ra mount Camera head with lens lens ring and cables Smart guide Camera adapter ring Extension block and cylinder Various screws and hex keys Power supply Camera head with lens lens ring and cables Smart guide Camera adapter ring Extension block and cylinder ZAP STAND ZAP STAND UPGRADE KIT ZAP WITH Ø76 ADAPTER KIT ...

Страница 4: ... and ZAP INLINE manual Then assemble the base and rod with four M5x16 countersunk screws bag 1 and place the assembled stand on a sturdy table Attach the camera mount to the rod with the three M4x12 screws bag 1 OPTION 1 WITH EXTENDED CAMERA ARM Mount the extension block on the camera head by using the four M4x60 screws bag 2 Mount the extension cylinder to the extension block with the two M4x10 s...

Страница 5: ... camera in the camera mount and tighten the front and side finger screws bag 3 OPTION 2 WITH CAMERA DIRECTLY IN CAMERA HOLDER Unscrew the lens with your hands and remove the lens ring by unscrewing the three M3x6 screws Screw the camera adapter ring on the camera with the three countersunk M3x10 screws bag 4 For correct assembly the laser label on the adapter ring must be placed next to the laser ...

Страница 6: ...efore tightening the front finger screw bag 3 Secure the cable bundle with the two P clips bag 3 If you re using a US model make sure to unmount lens protection cap when using this microscope ASSEMBLING ZAP STAND OR ZAP STAND UPGRADE KIT 3 3 ...

Страница 7: ...r current setup using the manufacturer s manual 6 ASSEMBLING ZAP WITH Ø76 ADAPTER KIT OPTION 1 WITH EXTENDED CAMERA ARM Then mount the extension block on the camera head by using the four M4x60 screws bag 1 Mount the extension cylinder to the extension block with the two M4x10 screws bag 1 ...

Страница 8: ...sembly the laser label on the adapter ring must be placed next to the laser itself Then rescrew the lens on the camera ASSEMBLING ZAP WITH Ø76 ADAPTER KIT 2 3 OPTION 2 CAMERA DIRECTLY IN HOLDER Place the extended camera arm with mounted camera in the camera mount and tighten the front and side finger screws bag 2 IMPORTANT Laser beam attenuator Unmount lens protection cap when using this microscop...

Страница 9: ...6 ADAPTER KIT 3 3 Place the extended camera arm with mounted camera in the camera mount and tighten the front and side finger screws bag 2 IMPORTANT Laser beam attenuator Unmount lens protection cap when using this microscope ...

Страница 10: ... power supply provided by TAGARNO Brand Model Mean Well GEM18I12 Protection class II Input 100 240V 0 45 0 2A 50 60Hz Overvoltage category ll Output 12V 1 5A Max 5 ESD Ring Terminal Connector ESD Grounding point 3000mm ESD Grounding point Connect HDMI cable 1 USB 3 0 cable 2 control box or foot switch 3 power supply 4 and ESD Electro Static Discharge 5 WARNING Do not use force when inserting the c...

Страница 11: ...h connect the cable to the foot switch and then connect the footswitch and control box Connect the power supply cable to the power supply battery and then to a electrical outlet Connect grounding cable to grounding point to ensure ESD of the microscope Connect USB 3 0 cable from the microscope to a PC ...

Страница 12: ...the power on 3 Keep holding the center button down for 25 seconds 4 Release the center button and turn the power off 5 Turn the power back on and the microscope is set to factory setting 1080P60 Turn off on Laser on off WARNING Laser radiation when turned on Zoom in Zoom out Push both plus and minus to change from auto focus to manual focus You can now adjust the focus manually by using the plus m...

Страница 13: ... plus and minus to change from auto focus to manual focus You can now adjust the focus manually by using the plus minus buttons When in focus mode If you wish to focus on something close If you wish to focus on something further away Foot switch for TAGARNO functions OPERATION 2 3 ...

Страница 14: ...format by pressing the center button and use arrow keys left right to select between 6 different formats 6 Select OSD presets You need to choose between respectively ALWAYS ON OFF or TIMEOUT using the left right arrows in order to have the OSD Menu displayed continuously never or for 3 seconds at the time Continue by pressing the down arrow 7 Color boost The OSD menu gives you the possibility to c...

Страница 15: ... 59 94 50Hz Camera zoom 30x optical Autofocus Yes Auto Monitor Detect No Power requirements Line voltage 100 240V 10 Line frequency 50 60Hz AC current consumption typ 70mA 100V 30mA 240V DC current consumption typ 0 5A 12V Environmental conditions Temperature Storage 5 to 60o C 23 to 140o F Operation 5 to 40o C 41 to 104o F Humidity range Storage 20 to 90 RH non condensing Operation 20 to 80 RH no...

Страница 16: ...s ring TAGARNO Ring light IR Glass table XKEY Control Box TAGARNO Ring light UV Cleaning kit Foot switch Backlight kit Circular tilting table FHD monitors per request Coaxial light FIELD OF VIEW MAXIMUM MINIMUM Lens X Direction Y Direction X Direction Y Direction 3 409 mm 16 10 230 10 mm 9 06 13 40 mm 0 53 7 54 mm 0 02 4 290 mm 11 42 163 10 mm 6 42 10 50 mm 0 41 5 91 mm 0 02 5 245 mm 9 65 137 80 m...

Страница 17: ...M with 2GB RAM Monitor resolution 1920x1080 recommended FULL HD 1080P 29 97 25HZ AND HD 720P 59 94 50 30 25HZ OS Desktop Laptop Windows 7 8 or 10 with DirectX 11 Memory 4GB CPU Intel Core Dual 3 0GHz or i3 2 4GHz 4th generation named 4xxx or newer Intel Core i3 3 2GHz 4th generation named 4xxx or newer Hard Disc Space 1GB required SSD type recommended Connections USB 3 0 xHCI host controller Intel...

Страница 18: ...ts that the product will correspond with the specification at the time of delivery and will be free from defects in material and workmanship for a period of 24 months 2 years from date of invoice from Supplier This warranty covers a The replacement of defective parts b All labour costs to exchange defective parts in product c Complete functions test of product before return to end user s site d Re...

Страница 19: ...camera unit and a 12V power supply is in compliance with the following European directives 2006 25 EU 2014 30 EU 2014 35 EU By conforming to the following harmonized standards IEC 60825 1 2014 EN 61326 1 2013 IEC 61010 1 2010 THE DECLARATION IS ISSUED BY MANUFACTURER TAGARNO A S Class 2 Class B Basic Electromagnetic Environment IECEE CB Scheme Ref Certif No NO104184 Artificial Optical Radiation El...

Страница 20: ...articular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an out...

Отзывы: