Disconnecting Battery
Open battery
compartment
and disconnect
battery as shown.
Plug battery
back in, controller
should beep twice
when initially
plugged back in
to the battery.
Controller
To power On/Off AffloVest, press and hold the (Standby) button on the controller.
After a brief startup period, the controller will display the main treatment menu.
Standby button is on top left of controller, press and hold a minimum
of two (2) seconds until the screen, like to the right, appears.
Treatments may be started at any time by pressing the
(Play/Pause)
button from the main treatment menu. Pressing the following buttons will
update the treatment whether AffloVest is running or not:
P
= Percussion Mode, all motors operate in a pulsed fashion.
V
= Vibration Mode, all motors operate continuously.
D
= Drainage Mode, motors operate in a pre-programmed
sequential fashion.
GO
= Automatically starts the Go Mode, the system will perform the
following treatment sequentially (by default):
Percussion Mode, 10 minute duration at medium intensity (followed by a 10 second pause).
Vibration Mode, 10 minute duration at medium intensity (followed by a 10 second pause).
Drainage Mode, 10 minute duration at medium intensity.
CONTROLLER ICON DEFINITIONS
P
Percussion Mode (motors pulse ON / OFF)
V
Vibration Mode (motors all ON)
D
Drainage Mode (motors ON in sequence)
High Intensity
Medium Intensity
Low Intensity
4