Model 1700 / 1750 Rotary Platform Abraser
18
Operating Instructions ver. 1.0
p/n 135505
The following steps are intended to assist in
establishing a test procedure for conducting wear
tests of rigid or flexible materials.
1. Mount the selected wheels on their respective
flanged holders.
IMPORTANT: Prior to testing, ensure the wheels
have been resurfaced, if necessary.
2. Measure and record specific parameters of the
unabraded test specimen, if necessary.
3. Secure specimen to Abraser specimen holder.
4. Select and mount the auxiliary weights, if
necessary.
5. Adjust the vacuum nozzle gap.
6. Using the Abraser touchscreen, enter the
appropriate test parameters then press the
“OK” button.
7. Lower the abrading heads onto the surface of
the test specimen.
8. Press START button on Abraser keypad to
commence testing.
9. To interrupt a test, press the PAUSE button. To
stop or cancel a test, press the STOP button.
NOTE:
If the STOP button is pressed before the
programmed number of cycles is reached, a screen
prompt will ask if you are sure. If you wish to continue
testing, select NO. Selecting YES will abort the test.