TMA Q Series Getting Started Guide
39
Probe Type
ContactDiameter
Pressure Exerted by
Types of Tests Yielded
mm (in.)
0.01 N Load
Flexure
5.08 (0.2)
0.49 kPa
Deflection temperature
Flexibility
Dilatometer
6.07 (0.239)
0.34 kPa
Expansion coefficient
Parallel Plate
9.52 (0.375)
0.14 kPa
Viscosity-temperature
Viscosity-time
Wall shear rate
Hemispherical
~2.54 (0.1)
~1.9 kPa
Softening point
NOTE: Refer to online help for details regarding the probes.
Zeroing the Auto Measure System
This procedure is used to initialize the auto-length measure system. It should be performed before each
experiment to ensure accurate sample length measurements. To zero the TMA auto measure system, simply
press the ZERO LENGTH key on the touch screen.