TPF / TFB Series Floor Scale Service Manual
- 14 -
Remove the leveling foot from the defective load cell and reinstall it in the
new load cell.
Tie the string to the new load cell cable and carefully pull the new cable
through the platform structure to the junction box opening. Coil any excess
cable and store it within the platform.
Bolt the new load cell and spacer plate to the platform. Apply an anti-seize
compound such as “Never –Seez” to the threads of the mounting screws
and tighten them with a calibrated torque.
Thread the load cell cable through the connector of the junction box.
Wire the new load cell cable to the proper summing connector.
Level the scale with the leveling feet.
Reattach the instrument cable and power up the scale terminal.
Perform a corner test and recalibrate the scale.