Load the first calibration mass weight on the platform (mass weight should
be1/3 of the max capacity) and wait few seconds for display stability.
Then press
, display will be shown
Load the second calibration mass weight on the platform (mass weight
should be2/3 of the max capacity) and wait few seconds for display
stability.
Then press
,display will be shown
Load the third calibration mass weight on the platform (mass weight should
be3/3 of the max capacity) and wait few seconds for display stability.
Then press
,display will be shown
After the calibration the display will start a self test. Remove the load from
platform during the test. Display will come to weighing mode automatically.
Load 2
Load 3
pass