During the working hours on the site, if an immediate communication of the recorded
values is needed
•
Set-up and activate the SMS feature (for the Alarm notification), and test it with the
dedicated button.
•
Check/insert the data network settings.
For more information, please refer to chapter 4.3.5.
STEP 3. Acquisition parameters
•
Choose the desired filter for the velocity sensor.
•
Choose the sampling rate. We suggest to select a sampling rate which does not exceed
10 times the maximum frequencies expected in the measurement.
For further details, please refer to chapter 4.4.2.
Figure 5.1.3. Settings of acquisition parameters.
STEP 4. Trigger recording
•
Choose the pre- and post- event duration.
•
Choose the max length.
•
Choose the trigger combination. We suggest to trigger when the threshold on one of the
three axes is exceeded (X or Y or Z)
•
We suggest not to activate the trigger increment
•
We suggest to choose the trigger levels after a site evaluation.
For further details, please refer to chapter 4.1.2.