Distal Radius Plate Instrument and Implant Set Surgical Technique DePuy Synthes 8
Apply plate
Place the plate on the distal radius and
determine which holes will be used for
fixation. The fracture pattern and clinical
situation will determine the specific order
of fixation (proximal leg or distal arm).
5
Secure distal arms
6
1.8 mm Threaded
Drill Guide
312.92
6A
: To secure the plate with 1.8 mm buttress
pins,
screw a 1.8 mm threaded drill guide into
the threaded plate hole until seated.
Note:
The threaded drill guide must be used to ensure
the proper drilling angle. Otherwise, the buttress pins
may not thread into the plate holes.
Determine whether 1.8 mm buttress pins
or 2.4 mm self-tapping cortex screws will
be used for fixation. A combination of both
implants may be used.