SYNRAD OEM v30 Operator’s Manual Version 2
Troubleshooting
Table 5-13
PWM Sense/Control Board fault
LED Indicator
LED Status
Output Signal
RJ45 Output Status
LASE (if tickle applied)
Off
n/a
– –
LASE (if PWM applied)
Off
n/a
– –
RDY Flashing
1
Laser OK
Low
Temperature OK High
PWR Flashing
1
Voltage OK
High
1 See the Laser fault indications section for details.
Laser fault indications
OEM v30 lasers have the ability to indicate five specific fault conditions. In the event of certain faults, the
RDY LED, PWR LED, or both will blink an error code, pause 1/2 second, and then repeat the code. This
sequence continues until the fault is corrected and the laser is reset by cycling DC power to the laser.
Table below lists fault codes, the corresponding fault condition, and describes possible corrective ac-
tions.
Table 5-14
Laser fault codes
LED
# of Blinks
Fault Condition
Corrective Action in Field
RDY
Continuous
Over Temperature fault
Cool laser and then cycle DC power
PWR
1 blink
Under Voltage fault
Verify 30 V (under load) and cycle DC power
PWR
2 blinks
Over Voltage fault
Verify 30 V (under load) and cycle DC power
PWR
3 blinks
DC Sense fault
Contact SYNRAD Customer Service
RDY/PWR Continuous
PWM Sense/Ctrl Bd fault
Contact SYNRAD Customer Service
Resetting faults
Remote interlock condition
A remote interlock condition occurs when the Laser Enable input opens (the RDY indicator turns off).
On OEM lasers, remote interlock conditions are not latched. Re-establish the Laser Enable signal input
(RDY LED changes from off to yellow) and begin lasing immediately.
Over temperature fault
Over temperature faults occur when thermal limits in the laser are exceeded. This fault is indicated by the
RDY indicator flashing continuously). To reset an over temperature fault, lower coolant temperature below
22 °C (water-cooled) or 40 °C (air-/fan-cooled) and then cycle DC power. When the RDY LED illumi-
nates without flashing, lasing is enabled.