7. Troubleshooting and Contact Information
NuGenius/
N
User manual
29
Chapter Seven - Troubleshooting and Contact
Information
7.1
Troubleshooting
No power to the darkroom
Check connection of main power cord to main power port on the rear of the
NuGenius/N.
Try another power socket within lab.
Transilluminator will not turn on
Check power cord by sliding Transilluminator fully out from cabinet. If loose
push back in.
If still not on, remove power cord and attach another one plugged in
elsewhere. If Transilluminator comes on there is an electrical supply problem
within the NuGenius/N. If it still does not come on it is likely the
Transilluminator has failed. Contact Syngene.
NOTE: Please take UV
precautions.
Print button is enabled but nothing happens when you select it
Check the correct printer has been selected.
If more than one printer is present, the system may be unable to distinguish
between them. Disconnect one and then continue.
Unable to save images with USB present
Check that the options on the “Image storage” screen are not disabled. If
they are, to re-enable them eject the USB and reinsert it. This should re-
enable the options and you will be able to save your image.