Poor Linearity
Usually a
mechanical
problem
Electronic
Problems
Re-check repeatability
*Inspection of the flexures for damage or
loose hardware may be required
*Use the Linearity Function in the service
menu to reset linearity
*A problem in the analogue circuit board or
power supplies can cause poor linearity.
Make sure all mechanical problems have
been eliminated first
OTHER PROBLEMS:
Cannot calibrate
Zero shifted more
than allowed
Calibration
timeout
*Check all flexures for damage
*Reset dealer calibration
*Verify linearity and repeatability
*The balance may be unstable. Verify
stability as above.
Calibration weight motor
does not stop
*Check the cables to the motor, try
plugging the balance into the power again
*Look for friction in the calibration weight
movement
*Check the opto-coupler that controls the
motor position.
RS-232 not working
Doesn’t print
Check parameters match the device
connected
Verify cable is correct
*RS-232 circuits damaged
Display dark, keys beep Display contrast
poor
Cable unplugged
or damaged
*Check the cables to the display
*Replace the display which could be
damaged
*To be carried out by authorised technicians only.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание 10001
Страница 2: ... 2014 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 4: ...2 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 11: ... 2014 9 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 12: ... 2014 10 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 49: ... 2014 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 53: ... 2014 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...