5120A/5115A Operations and Maintenance Manual
67
7.3 The Network Data Port
The network data port outputs phase difference data while the test set is in the
Collecting
state. The
phase differences are in units of cycles of the Input signal. Phase differences are output as floating
point numbers, one number per line, either 1, 10, 100, or 1000 samples each second. The default
rate is 1000 samples per second, to change the rate use the
set phaserate
7, Setting Data Port Output Rate”
set print local
Print new jobs to local (USB) printer.
set print remote <host> <port>
Print new jobs to network printer.
set referencefreq <auto|freq>
Selects automatic, or specified manual reference frequency used
by frequency counter.
set tau0 <tau0value>
Sets tau0 value in seconds: 0.001, 0.01, 0.1 or 1
set timeconstant <value>
Sets length of running average time constant for spectrum and
Allan deviation
set timeformat <24/12>
Selects time format for printout time stamps.
set title <title>
Sets user specified title (quoted string up to 34 characters)
show adev
Returns snapshot of Allan deviation tables.
show date
Returns current date and time in selected format.
show dateformat
Returns selected printout date stamp format.
show fcounter
Returns snapshot of frequency counter table.
show freqdiff
Returns snapshot of frequency difference strip chart.
show inputs
Returns input frequency and amplitude on input signal ports.
show ipn
Returns integrated phase noise table.
show mac
Returns MAC (ethernet) address of the test set.
show phasediff
Returns snapshot of phase difference strip chart.
show phaserate
Returns current phaserate value.
show printformats
Returns acceptable values for
set print format
command.
show printoptions
Returns current printer settings.
show screens
Returns name of screens that can be printed.
show spectrum
Returns snapshot of spectrum table.
show spurs
Returns snapshot of spur table.
show state
Returns current collecting state.
show tau0
Returns current tau0 value.
show timeformat
Returns selected printout time stamp format.
show title
Returns current user specified title.
show version
Returns model, software and hardware version information.
start
Begin data collection
stop
Stop data collection.
Table 4: Command Summary
Command
Synopsis
Содержание 5115A
Страница 6: ...iv Contents Index 97 ...
Страница 12: ...6 1 Introduction ...
Страница 16: ...10 2 Setting Up a PNTS ...
Страница 26: ...20 3 The Screen and Basic Concepts ...
Страница 82: ...76 9 Maintaining a PNTS ...
Страница 92: ...86 Appendix A Specifications ...
Страница 98: ...92 Appendix B Theory of Operation ...
Страница 100: ...94 ...
Страница 102: ...96 ...
Страница 106: ...100 ...