SWP-CA102 Multi-Functions Calibrator Operating Instruction

 

 

The “Frequency” output and measuring: 

 

 

 

 

3.4 Parameters setting 

Press ENT will go into parameter setting, ESC will be back to main screen. The parameters as shown in below: 

 

 
 
 
 
 
 

 
The compensation value can be key-in or modified. Press ENT to confirm and ESC back the main screen after 

setting.

 

3.4.1 General parameters setting  

Parameter 

Range 

Default 

Setting 

Description 

Input 

compensation 

Manual or Auto 

Manual 

Cold-junction  compensation  modes  in  T.C  measuring: 

Manual

 compensation parameter setting; 

Auto

 compensation setting which CA102 detected by T.C 

Cold-Junction temperature.   

Temperature 

Compensation 

00.0

~99.9

 

00.0 

Cold-junctions 

compensation 

value 

of 

T.C 

input 

measuring in manual mode.   

Output 

Compensation 

Manual or Auto 

Manual 

Cold-junctions compensation value for T.C output manual 

setting.   

Temperature 

Compensation 

00.0

~99.9

 

00.0 

T.C  input  measuring  cold-junctions  compensation  for 

manual setting.  

Contrast 

0~4 

02 

Screen contract setting 

password 

0000~9999 

0000 

For into the second level parameter setting page 

3.4.2 Second level (Calibration) parameter setting

For professional engineer only

 

Key in the password, press ENT,         then move to second level parameter setting 

Parameter 

Range 

Default 

setting 

description 

Password  

0000~9999 

0000 

Password change or verify 

DCVZero 

-9999~+9999 

0000 

DC V Offset for cal 

DCVProp 

0.000~9.999 

1.000 

DC V proportion for cal 

DCmAZero 

-9999~+9999 

0000 

DC mA offset for cal 

DCmA Prop 

0.000~9.999 

1.000 

DC mA proportion for cal 

DCmVZero 

-9999~+9999 

0000 

DC mV Offset for cal 

DCmVProp 

0.000~9.999 

1.000 

DC mV proportion for cal 

RTD1Zero 

-9999~+9999 

0000 

RTD1 (Cu50) offset for cal 

RTD1Propo 

0.000~9.999 

1.000 

RTD1 (Cu50) proportion for cal 

RTD2Zero 

-9999~+9999 

0000 

RTD2 (Pt100) offset for cal 

 

OUT 

 

0-1000Hz 

 

0000Hz 

 

 

IN 

 

0000.0H

 

 

 

 

 

 

In Comp 

Auto 

Out Comp 

00.0

 

 

Comp Temp 

Comp Temp 

Auto 

00.0

 

Содержание SWP-CA102

Страница 1: ...SWP CA102 Multi Functions Calibrator Operating Instruction 1 SWP CA102 Multi Functions Calibrator Operating Instruction S W P 2 0 0 9 S e r i e s...

Страница 2: ...S 90 standard query value is available T C Measure and source cold junctions compensation in auto and manual modes Battery continually usage life is about 5 7 hours per charging Key sound and backligh...

Страница 3: ...DCV DCmA DCmV RTD TC F The source sequence is DCV DCmA DCmV R TC F Range In measure and source further select the types of RTD TC F RTD Cu50 Pt100 TC S K E T J B R N W325 F 0 1000Hz 0 5 00KHz Zero Ou...

Страница 4: ...the number directly Output Source resistance or Measure resistance the screen is shown as above In the measuring there are some types to select Measure Range it will switch to the follow the sequence...

Страница 5: ...put manual setting Temperature Compensation 00 0 99 9 00 0 T C input measuring cold junctions compensation for manual setting Contrast 0 4 02 Screen contract setting password 0000 9999 0000 For into t...

Страница 6: ...cal TC NZero 9999 9999 0000 N type TC offset for cal TC NProp 0 000 9 999 1 000 N type TC proportion for cal TC WZero 9999 9999 0000 W type TC offset for cal TC WProp 0 000 9 999 1 000 W type TC propo...

Страница 7: ...t is necessary to on hold the instrument at least 5 minutes to stable the device if the compensation mode is set to AUTO and the instrument has moved to a new temperature environment For T C signal ou...

Отзывы: