SWP-CA102 Multi-Functions Calibrator Operating Instruction
5
The “Frequency” output and measuring:
3.4 Parameters setting
Press ENT will go into parameter setting, ESC will be back to main screen. The parameters as shown in below:
The compensation value can be key-in or modified. Press ENT to confirm and ESC back the main screen after
setting.
3.4.1 General parameters setting
Parameter
Range
Default
Setting
Description
Input
compensation
Manual or Auto
Manual
Cold-junction compensation modes in T.C measuring:
Manual
compensation parameter setting;
Auto
compensation setting which CA102 detected by T.C
Cold-Junction temperature.
Temperature
Compensation
00.0
℃
~99.9
℃
00.0
Cold-junctions
compensation
value
of
T.C
input
measuring in manual mode.
Output
Compensation
Manual or Auto
Manual
Cold-junctions compensation value for T.C output manual
setting.
Temperature
Compensation
00.0
℃
~99.9
℃
00.0
T.C input measuring cold-junctions compensation for
manual setting.
Contrast
0~4
02
Screen contract setting
password
0000~9999
0000
For into the second level parameter setting page
3.4.2 Second level (Calibration) parameter setting
(
For professional engineer only
)
Key in the password, press ENT, then move to second level parameter setting
Parameter
Range
Default
setting
description
Password
0000~9999
0000
Password change or verify
DCVZero
-9999~+9999
0000
DC V Offset for cal
DCVProp
0.000~9.999
1.000
DC V proportion for cal
DCmAZero
-9999~+9999
0000
DC mA offset for cal
DCmA Prop
0.000~9.999
1.000
DC mA proportion for cal
DCmVZero
-9999~+9999
0000
DC mV Offset for cal
DCmVProp
0.000~9.999
1.000
DC mV proportion for cal
RTD1Zero
-9999~+9999
0000
RTD1 (Cu50) offset for cal
RTD1Propo
0.000~9.999
1.000
RTD1 (Cu50) proportion for cal
RTD2Zero
-9999~+9999
0000
RTD2 (Pt100) offset for cal
OUT
F
0-1000Hz
0000Hz
IN
F
0000.0H
z
In Comp
Auto
Out Comp
00.0
℃
Comp Temp
Comp Temp
Auto
00.0
℃