5、Select “OK” to do auto-calibratio.Calibrator will measure the color of screen and calibrate the
screen automatically,please pay attention to the progress bar.
6、Press “INFO” to stop calibration.When the progress bar reach to 100%,auto calibration finished.
Press “INFO”to exit menu after auto-calibration,monitor will recover to normal.
7、The screen will show the Delta value before and after calibration,shown as below:
*4 Measurement
Monitor have calibrated in factory already but also need to re-calibrate after a period of time
using.Before re-calibrating,“Measurement” function will show Delta value of last time calibration,
user can decide to calibrate or not from comparing the Delta value.
15
Auto Calibration
Progress 23%
Please wait...
Select INFO to Cancel!
INFO
:
Quit
Auto Calibration
Please make sure the monitor is in
dark room,and the sensor probe is
put in the center marker and close
to the screen.
Select MENU to start!
MENU:OK INFO:Quit
Auto Calibration
Please make sure the monitor is in
dark room,and the sensor probe is
put in the center marker and close
to the screen.
Select MENU to start!
MENU:OK INFO:Quit
R
R75
Mg
Mg75
B
B75
Cy
Cy75
G
G75
YI
YI75
W
Delta = 0.9
Before Calibration
After Calibration
R
R75
Mg
Mg75
B
B75
Cy
Cy75
G
G75
YI
YI75
W
Delta = 2.3
CIE and Delta E
before calibration
CIE and Delta E
after calibration
R
R75
Mg
Mg75
B
B75
Cy
Cy75
G
G75
YI
YI75
W
Delta = 0.9
Last Calibration
New Measured
R
R75
Mg
Mg75
B
B75
Cy
Cy75
G
G75
YI
YI75
W
Delta = 2.3
CIE and Delta E
result of last calibration
CIE and Delta E
result of new measured
Connect calibrate probe with monitor and
put it on the right position,Select “Auto
Calibration”-“Measurement”,Monitor will
create part of color and finish testing within
30 seconds.Testing result shown as below: