background image

21

VIII. Troubleshooting table

As you use your microscope, you may occasionally experience a problem.  
The troubleshooting table below contains the majority of frequently 
encountered problems and the possible causes.

Optical 

 

e

s

u

a

C

 

e

l

b

i

s

s

o

P

Problem

Vignetting or uneven brightness 

in the field of view or field of view 

only partially visible   

Lamp not installed properly 
Condenser not mounted correctly 
Condenser is set too low 
Aperture diaphragm closed too far 
Revolving nosepiece not clicked into position 
Filter not in placed in properly 

Blur image or dirt in the field of view 

Aperture diaphragm closed too far 
Dust or dirt on specimen surface 
Dust or dirt on field lens, filter, condenser or 

eyepiece 
No cover glass 
Too thick or thin cover glass 
Immersion oil not used with oil immersion lens 
Air bubbles in immersion oil 
Specified immersion oil not used 
Immersion oil on dry objective 
Greasy residue on eye lens  
Incorrect illumination 

Uneven focus 

Specimen holder not fixed securely on stage 
Specimen not secured in position 
Specimen tilted on stage surface 

Image tinged yellow 

Lamp voltage is set too low 
Blue filter is not being used 

Focusing is not possible with high 

magnification objectives 

 

Slide is upside down 
Cover glass is too thick 

Содержание Stellar 1 Pro-T

Страница 1: ...SHOP SWIFTOPTICAL COM STELLAR 1 SERIES Biological Microscope Instruction Manual M TEC LIMITED...

Страница 2: ...irements of modern research techniques and testing methods This involves modification to the mechanical structure and optical design of our instruments Therefore all descriptions and illustrations in...

Страница 3: ...ness of this small glass piece is now standardized at 0 17 mm for most applications Diaphragm Condenser A diaphragm which controls the effective size of the condenser aperture A synonym for the conden...

Страница 4: ...id is usually required by objectives of 3 mm focal length or less Resolving Power A measure of an optical system s ability to produce an image which separates two points or parallel lines on the objec...

Страница 5: ...erpupillary distance adjustment 3 Diopter adjustment 4 Coarse and fine focusing 9 Use of Darkfield 5 Use of Parfocal guide line 6 Coarse focus torque adjustment 7 Stage upper limit stop adjustment 8 U...

Страница 6: ...be Objective Color Ring Color Coded Aperture Diaphragm Transmitted Light Collector Specimen Holder lever Eyepiece Tube Clamp Screw Power Adapter Compartment Brightness Indicator Revolving Nosepiece Co...

Страница 7: ...o 106 kPa Avoid frost dew percolating water and rain III Assembling the microscope 1 Verifying input voltage The 5V microscope input can works with a broad range of USB power sources However always us...

Страница 8: ...ation using natural light in a location where power supply is not available For this purpose it is necessary to remove the transmitted light collector Raise the stage up to the top position Rotate the...

Страница 9: ...of the condenser is to provide full illumination to the specimen plane and to enhance the resolution and contrast of the object being viewed 6 Eyepiece tube Eyepiece tube is preinstalled in microscop...

Страница 10: ...of microscope base will light up When the light intensity control wheel is turned anti clockwise to high brightness position the light intensity will increase Automatic power off function A safety an...

Страница 11: ...osepiece to 10x agnification Using the right eye only and viewing through the right hand eyepiece adjust the focus with the microscope coarse or fine adjustment until the image of the specimen is at i...

Страница 12: ...further than their limit Coarse Focus Torque Adjustment Fine Focus Coarse Focus 5 Use of Parfocal guide line The parfocal guide lines are in the stage holder and microscope body to assist in quick ima...

Страница 13: ...can move the stage regardless of the limit but will only lower the stage Lower the stage by coarse focus knob Stage Upper Limit Stop 8 Use of condenser aperture diaphragm Critical illumination relies...

Страница 14: ...the corresponding range of the objective color ring For example red range for 4x Adjust the aperture diaphragm slightly to get the sharpest image according to specimens 4X 10X 40X 100X 4X 10X 40X 100...

Страница 15: ...scope Bring the centering telescope into focus so that the black ring of objective phase plate and segmented phase annulus of phase slide are in focus Use the knurled screws to center the black ring o...

Страница 16: ...ur computer If you would like to mount an eyepiece or eyepiece camera not included directly on it screw the provided eyepiece ring into the top of the adapter and insert the eyepiece or eyepiece camer...

Страница 17: ...roof device For the same total magnification select a combination of the highest possible objective magnification and lowest possible projection lens magnification to achieve the utmost image definiti...

Страница 18: ...iped off the objective with a lens cleaning tissue and the residual film removed with soft cloth moistened with petroleum benzene or absolute alcohol Locate the field of interest with a lower magnific...

Страница 19: ...of alcohol and ether ratio alcohol 3 and ether 7 to remove immersion oil from objective lenses Because the mixture of alcohol and ether ratio alcohol 3 and ether 7 is highly flammable be careful hand...

Страница 20: ...o not use the microscope near water E Bulb replacement The lamp and the lamphouse become very hot during and after a period of operation Risk of burn Do not touch the lamp during or immediately after...

Страница 21: ...torage Insert the power adapter into power adapter compartment on the rear of microscope when it is not in use Power cord storage When it s not in use insert the USB end of the power cord into power a...

Страница 22: ...placed in properly Blur image or dirt in the field of view Aperture diaphragm closed too far Dust or dirt on specimen surface Dust or dirt on field lens filter condenser or eyepiece No cover glass To...

Страница 23: ...not adjusted Eye strain or fatigue Interpupillary distance not adjusted Diopter adjustment not made Field of view of left and right eyepiece differ Inadequate illumination Electrical Lamp does not lig...

Страница 24: ...Video components Does not cover light bulbs batteries fuses or electrical cords All warranties start from the original date of purchase Swift provides the repair or replacement of warrantied parts fo...

Страница 25: ...0 50 30 inclined Binocular 30 inclined Trinocular light distribution 50 50 30 inclined Interpupillary Distance Eyepieces Nosepiece Objectives Magni cations Stage Focus Illumination Microscope Power In...

Страница 26: ...llar 1 T Ph Binocular Finite Trinocular Finite WF10X WF16X WF25X Quadruple Head Tube Eyepiece Nosepiece Objective Stage SA 4X 10X 40X 100X SA 4X 10X 100X S Plan Ph40X SA Plan 4X 10X 40X 100X SA 4X 100...

Страница 27: ...epiece Objective Stage S Plan In nity 4X 10X 40X 100X S Plan In nity 4X SA Plan In nity 4X 10X 40X 100X 10X 100X Ph40X S Plan In nity 4X 100X Ph10X Ph40X Mechanical Stage Condenser Functional Accessar...

Страница 28: ...customersupport swiftoptical com M TEC XIAMEN CO LTD 3 A301 No 810 Fangshan Nan Road XiangAn Xiamen P R China Tel 86 592 569 8191 Mail customersupport swiftoptical com Website Shop swiftoptical com E...

Отзывы: