IMPORTANT:

 The working distances of the 40XRD and 100XRD objectives 

to the slide surface are very small and although the oil immersion 
objectives are sealed to prevent oil contamination, it is a good practice 
to avoid dragging these objectives through an oiled slide. The 100XRD oil 
immersion lens on Swift microscopes has a spring-loaded end to prevent 
cracking the cover slip upon its initial contact. Once this zone of safety is 
exceeded by moving the slide further toward the lens, a point can be 
reached where damage will occur. Always make a practice of frequently 
checking the position of the lens on the slide. Note that a 100XRD 
objective requires an iris diaphragm for brightfield oil immersion 
microscopy. 
 
 
 
 
 
 
 

 

 

 

HOW TO USE THE POINTMASTER

® 

EYEPIECE RETICLE 

 

Swift's patented POINTMASTER® eyepiece reticle, which is installed in 
the eyepiece of the M3500 microscope, enables the user to easily 
measure the size of the specimen. 
 
I.  The numbers on the chart indicate the actual size of the  
  

POINTMASTER® scale in millimeters. For example, the length is 

  

1.0mm from the top of the arrow to the bottom; the total length of 

  

the scale is 8.0mm, etc. The thickness of all the horizontal lines is 

  

0.01mm. 

 
2.  To obtain the actual physical size of a specimen, divide the 

POINTMASTER® scale readings by the magnification of the objective 
lens that is in use. Divide the readings by 10 if the 10X objective is 
being used, 40 if the 40XRD objective is in use, or 100 if of 100XRD is 
in use. 

 
 
 

Содержание M3500 SERIES

Страница 1: ...CROSCOPE The Swift M3500 Series microscope is considered to be the most Student proof microscope on the market It is an instrument of optical and mechanical precision and will perform satisfactorily w...

Страница 2: ...age drift CONDENSER the condenser is mounted in the stage and it is used in conjunction with the iris diaphragm The function of the condenser is to provide full illumination to the specimen plane and...

Страница 3: ...ide in place allowing the user to move the slide on any x y axis through the manipulation of two control knobs or a co axial control mechanism NOSEPIECE the revolving turret that holds the objective l...

Страница 4: ...COVER GLASS Thin glass cut in circles rectangles or squares for covering the specimen usually a thickness of 0 15 to 0 I7mm The majority of specimens should be protected by a cover glass and must be c...

Страница 5: ...esolving power is directly proportional to the numerical aperture of the optical system WORKING DISTANCE the distance from the lens of the objective to the cover slip on the slide when the specimen is...

Страница 6: ...ification of 400X or lower However at a magnification of 1 000X and above refraction becomes problematic This problem is reduced significantly by placing a thin layer of very clear viscous oil between...

Страница 7: ...100XRD objective requires an iris diaphragm for brightfield oil immersion microscopy HOW TO USE THE POINTMASTER EYEPIECE RETICLE Swift s patented POINTMASTER eyepiece reticle which is installed in the...

Страница 8: ...ware must first be installed on a computer The minimum computer requirements to use the camera is having an available USB 2 0 port Windows XP Vista or Mac OS X operating system installed on the comput...

Страница 9: ...e microscope is not in use CLEANING The front lens of the objectives particularly the 40XRD and 100XRD should be cleaned after use First brush with a soft camel hair brush or blown off with clean oil...

Страница 10: ...ay corrode your microscope Also be sure that the objectives are placed in the lowest possible position and the rotating head is turned inward and not protruding from the base Microscopes with mechanic...

Страница 11: ...s electrical cord from the electrical socket and remove any slides on the stage Use the small allen wrench 09mm that was included with the microscope to loosen the set screws that hold the black illu...

Страница 12: ...e enough Slowly open the diaphragm to increase the size of the opening allowing for additional illumination E PROBLEM Once the specimen is in focus it moves out of focus CORRECTION Gravity is causing...

Страница 13: ......

Отзывы: