AMI Turbitrace
Program List and Explanations
A-96.250.131 / 050220
59
Press the [Enter] key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val
-
ues will be reset.
3.4 Replace Absorber
If maintenance message E068 or error message E006 appears,
call this menu item and replace the humidity absorber. This will
make both messages disappear for 3 days. This time is needed for
the new humidity absorber to dry the housing.
3.5 Set Time
Adjust date and time.
4 Operation
4.1 Sensors
4.1.1
Filter time constant:
Used to damp noisy signals. The higher the fil
-
ter time constant, the slower the system reacts to changes of the
measured value. Range: 5–300 sec
4.1.2
Hold after cal:
To allow the instrument to stabilize again after cali
-
bration. During cal. and hold time the signal outputs are frozen,
alarms and limits are not active. Range: 0–6000 sec
4.2 Relay Contacts
See chapter 5 Installation.
4.3 Logger
The instrument is equipped with an internal logger. The logger data
can be copied to a PC with an USB stick if option USB interface is
installed.
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measured value, temperature, flow.
Alarm Relay:
Active or inactive
Relay 1 and 2:
Active or inactive
Input:
Open or closed.
Signal Output 1 and 2:
Actual current in mA.
Signal Output 3:
Actual current in mA (if option is installed)