24
A-96.250.321 / 191015
AMI Solicon4
Installation
3.4.4
Adjust the deltaT sensor
The deltaT flow sensor is factory calibrated at 20 °C (± 20 % accu-
racy).
The accuracy of the flow measurement depends on the ambient
temperature of the location where the deltaT sensor is installed. If
the ambient temperature is higher or lower than 20 °C, the deltaT
flow sensor can be adjusted. If an adjustment of the deltaT sensor
is necessary proceed as follows:
Run in
After installation let the sensor run-in for at least 1h.
Determine the
flow rate
1
Put the sample outlet of the instrument for 10 min. into a mea-
suring cup with a sufficient volume.
2
To get the flow rate in l/h, multiply the amount of water con-
tained in the measuring cup with factor 6.
The flow rate in l/h results from the multiplication of the
amount of water after 10 min by 6.
Adjust slope
1
Navigate to <Installation/Sensors/Flow>, choose <Slope> and
press [Enter].
2
If the calculated flow rate is higher than the displayed flow rate
increase the Slope value.
3
If the calculated flow rate is lower than the displayed flow rate
decrease the Slope value.
4
Press [Exit] and save with [Enter].
5
Compare the calculated flow rate with the displayed flow rate.
If the flow rates are roughly equal, the adjustment is finished.
Else repeat step 1 to 5.