A-96.250.211 / 171121
AMI Sodium P
Installation
31
3.4.3
Installing the pH Electrode
1
Carefully remove the protective cap [D] from the electrode tip.
Turn it clockwise only.
2
Rinse the electrode tip with clean water.
3
Slip union screw [E], and washer [F] on the electrode shaft [C].
4
Wet the O-ring [G] and slip it carefully over the electrode shaft.
5
Insert the electrode through the hole [H] in into the flow cell.
6
Tighten the union screw [E] hand-tight.
7
Remove the connector cap [A].
8
Screw the connector [B], of the cable marked with pH, onto the
pH electrode.
9
Keep the protective caps on a secure place for later use.
10
Connect the cable to the front end in the AMI Transmitter, see
A
B
C
D
E
F
G
H
Connector cap
Sensor plug
Sensor shaft
Protective cap
Union screw
Washer
O-ring
pH sensor hole
A
B
C
D
E
F
G
H