AMI Powercon
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1
Flow
:
None
Q-Flow
Select “Q-Flow” if the sample flow should be monitored and shown
on the display and when using a SWAN flow cell.
5.1.2
Sensor parameters
5.1.2.1
Cell Constant
: Enter the cell constant printed on the sensor label.
5.1.2.2
Temp. Corr
: Enter the temperature correction printed on the sensor
label.
5.1.2.3
Cable length
: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2.4
Meas. unit
: Select the measuring unit as
s/cm or as
s/m.
5.1.3
Temp. comp
:
5.1.3.1
Comp
.: Available compensation models are:
none
Coefficient
Neutral salts
High purity water
Strong acids
Strong bases
Ammonia, Eth.am.
Morpholine
5.1.4
Quality Assurance
:
5.1.4.1
Level
.: Choose the quality level according to your requirements.
0: Off; Quality Assurance is not active.
1: Trend (details see
Quality assurance level, p. 52
2: Standard (details see
Quality assurance level, p. 52
3: Crucial (details see
Quality assurance level, p. 52
4: User; edit user specific limits in menu 5.1.4.2 - 5.1.4.4