AMI Codes-II CC
Program List and Explanations
5.3.1.6
Alarm pH:
Only available if pH option is installed. Define the mea-
suring value, which should issue an alarm high respectively low.
5.3.1.6.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E039 is displayed in the mes-
sage list.
Range: 0–14 pH
5.3.1.6.26
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E040 is displayed in the message
list
Range: 0–14 pH
5.3.1.6.36
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value
Range: 0–14 pH
5.3.1.6.46
Delay
: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec