AMI INSPECTOR Hydrogen
Error List
A-96.250.741 / 050218
41
Error
Description
Corrective action
E001
Hydrogen Alarm high
– check process
– check programmed value,
see
5.3.1.1.1, p. 58
E002
Hydrogen Alarm low
– check process
– check programmed value,
see
5.3.1.1.22, p. 58
E003
Saturation Alarm high
– check process
– check programmed value,
see
5.3.1.4, p. 59
E004
Saturation Alarm low
– check process
– check programmed value,
see
5.3.1.4, p. 59
E007
Sample Temp. high
– check process
– check programmed value,
see
5.3.1.3.1, p. 59
E008
Sample Temp. low
– check process
– check programmed value,
see
5.3.1.3.22, p. 59
E009
Sample Flow high
– check sample flow
– check programmed value,
see
5.3.1.2.2, p. 59
E010
Sample Flow low
– establish sample flow
– clean instrument
– check programmed value,
see
5.3.1.2.32, p. 59
E011
Temp. shorted
– Check wiring of sensor
– Check sensor
E012
Temp. disconnected
– Check wiring of sensor
– Check sensor