AMI INSPECTOR Hydrogen
Maintenance
A-96.250.741 / 050218
29
6.3.2
Calibration
The hydrogen sensor of the AMI INSPECTOR Hydrogen is calibrat-
ed with pure hydrogen. To perform a calibration proceed as follows:
NOTICE:
Do not expose the sensor to 100 % Hydrogen longer
than 6 minutes.
1
Stop the sample flow at the main tap.
2
Navigate to menu <Maintenance> /<Calibration>.
3
Press [Enter] to start the calibration and follow the dialog on the
display.
A
B
C
D
E
Threaded sleeve
Hydrogen sensor
Quick-lock coupling
Hydrogen tube
Sample tube