Press button 1 to confirm the selection. The display shows “1.out” again. 

When pressing button 1 again, the device will display “1.Err” (preferred state of output 1).

Use button 2 or button 3 to set the desired initial state in case of an error.

Display

Kind of Output

Note

Off

Inactive in case of an error

Active in case of an error

Low- / High-side switch is opened in case of an error. 
Push-Pull output is low in case of an error.

On

Low- / High-side switch is closed in case of an error. 
Push-Pull-output is high in case of an error.

Press button 1 to confirm the selection. The display shows “1.Err” again. 

If you selected a 3-point controller, the level 2 settings must be the same as level 1 settings.
2.dEL” (delay of output 2), “2.out” (kind of output 2), “2.Err” (preferred state of output 2).

When pressing button 1 again, (only if you configured the device with min/max alarm), the device will display
A.out“ (type of alarm output).

Use button 2 or button 3 to select the type of alarm output.

Display

Kind of Alarm Output

Note

nPn

Low-Side 

NPN, open collector, switching GND

High-Side

PNP, open collector, swiUb

Push-Pull

Switching output is closed (connected to GND) as long there 
is no alarm condition and is opened if there is an alarm condition.

PnP

Pu.Pu

Switching output is high with no alarm condition and changes to 
low if there is an alarm condition.

Switching output is closed (is under voltage) as long there is 
no alarm condition and is opened if there is an alarm condition.

Press button 1 to confirm the selection. The display shows “A.out” again. 

Depending on the selected output function, you have to make the settings for switching or alarm points.
See description in section “switch points and alarm boundaries” for further information.

Note: The settings for the switching and alarm points can be made later in an extra menu (see Section 5).

Note:  The switching outputs are inverted when used as alarm outputs. This means as long there is no alarm condition, 

the switching output will be active. In case of an alarm condition, the output will become inactive.

Note:  When using the output function “min / max alarm, individual“ the setting for type of alarm output is used for both alarm

outputs.

Содержание PTI-PM

Страница 1: ...Remote Panel Mount Display Model PTI PM www swagelok com Operating Instructions...

Страница 2: ...ing outputs 7 3 4 1 Connection with configured low side switching output NPN output switching to GND 7 3 4 2 Connection with configured high side switching output PNP output switching to Uv 7 3 4 3 Co...

Страница 3: ...es 1 Use within the conditions specified in the Swagelok General Industrial Transducer catalog 2 Always disconnect the device from its power supply before opening it Do not touch any of the device s c...

Страница 4: ...d 3 will increase or decrease in value by one each time the button is touched If the button is held down the value will increase or decrease quickly 4 MS 13 PTI PM E Rev 0 6 04 WEL To configure the PT...

Страница 5: ...mA 8 and 7 6 and 7 0 V 0 V 0 V 3 3 V 3 3 V 10 V 1 V 1 V 1 V 4 V I 10 mA 10 V I 10 mA 20 V Note The contacts 3 5 and 7 are connected internally 3 2 Connection data Do not exceed the limit of the curren...

Страница 6: ...cer Sig 6 10 V 1 V 2 V Supply 9 to 28 V dc Transducer Uv Uv Sig 4 5 7 9 6 10 V 1 V 2 V 4 5 7 9 Supply 9 to 28 V dc 0 4 to 20 mA transducer Supply for transducer Uv Uv Uv Uv 4 5 9 Sig 4 5 7 9 Supply 9...

Страница 7: ...rm condition occurs Note In order to avoid unwanted or wrong switching processes connect the device s switching outputs after you have configured the device s switching outputs properly Connecting con...

Страница 8: ...the input signal SEnS and the measuring unit Unit all settings will be changed to factory default You have to set all the other settings including the settings for zero and span adjustment and switchi...

Страница 9: ...ay when a 0 mA 4 mA or 0 V input signal is attached Confirm the selected value by pressing button 1 The display shows di Lo again Press button 1 again The display will show di Hi Display High high dis...

Страница 10: ...2 point controller 2P 5 1 3 point controller digital 2 point controller 5 1 2 point controller with Min Max alarm digital 2 point controller 2P AL 5 2 Min Max alarm common Max alarm AL F1 5 3 Min Max...

Страница 11: ...isplay Kind of Alarm Output Note nPn Low Side NPN open collector switching GND High Side PNP open collector switching Ub Push Pull Switching output is closed connected to GND as long there is no alarm...

Страница 12: ...e will display 1 on turn on point of output 1 Use button 2 or button 3 to set the desired value The device s output 1 will turn ON Press button 1 to confirm your selection The display shows 1 on again...

Страница 13: ...on Note The unit of the value will be in sec The device will turn on the alarm after the minimum or the maximum alarm value was active for the delay time you have set Press button 1 to confirm the del...

Страница 14: ...ithout span adjustment the device would display 1020 with span adjustment of 102 Press button 1 to confirm the selection of the span adjustment The display shows SCAL again Example Connecting a 4 to 2...

Страница 15: ...ecessary Err 4 Values below display range Indicates that display value is below the valid display range of the device 1999 digit Possible causes Incorrect scale Remedies The error message will be rese...

Страница 16: ...Swagelok TM Swagelok Company 2004 Swagelok Company September 2004 R0 MS 13 PTI PM E...

Отзывы: