In-situ check with SV 111
1) Open the SV 111 calibrator lid and take out the SA111 adapter
Mount the SV 100A on the SA111.
2) Mount the SV 100A on the SA111 (x axis vertical).
Set up of SV 111
5) Start the calibrator using level of 1 m/s
2
at 16 Hz and wait until all
3 LEDs are green.
SV 100A User Guide
5
4) Enable Calibration menu in SV 100A by holding Right Arrow button.
Select the axis.
6) Press Enter on
SV 100A to start the
check procedure.
7) Repeat the procedure
in all 3 axes.