Svantek SV100A Скачать руководство пользователя страница 5

In-situ check with SV 111

1) Open the SV 111 calibrator lid and take out the SA111 adapter  
  Mount the SV 100A on the SA111.

2) Mount the SV 100A on the SA111 (x axis vertical).

Set up of SV 111

5) Start the calibrator using level of 1 m/s

2

 at 16 Hz and wait until  all  

  3 LEDs are green.

SV 100A User Guide

5

4) Enable Calibration menu in SV 100A by holding Right Arrow button.  
  Select the axis. 

6) Press Enter on  

 

  SV 100A to start the  
  check procedure. 

7) Repeat the procedure 
  in all 3 axes.

Содержание SV100A

Страница 1: ...SV100A Vibration Dosimeter User Guide...

Страница 2: ...s in the SV 100A automatically detect the presence of a user or vehicle driver which enables real daily exposure calculations for the period of time when the user is in contact with the vibrating surf...

Страница 3: ...lysis Veri cation of vibration sources Research and development SF 100A 3OCT 1 1 1 3 octave analysis Veri cation of vibration sources Research and development SF 100A WAV Time domain signal recording...

Страница 4: ...settings using panels on the right hand side 8 Disconnect the USB cable Instrument is ready to measure SV 100A User Guide 4 2 Connect the USB cable to PC Make sure the Supervisor software and USB dri...

Страница 5: ...100A on the SA111 x axis vertical Set up of SV 111 5 Start the calibrator using level of 1 m s2 at 16 Hz and wait until all 3 LEDs are green SV 100A User Guide 5 4 Enable Calibration menu in SV 100A b...

Страница 6: ...ndroid smarphone 3 Start the application The Assistant will ask you to enable Bluetooth and Localization services on your smartphone tablet 4 The Assistant will detect the SV100A automatically Install...

Страница 7: ...open file simply double click on it SV 100A User Guide 7 2 Select files you wish to download and click the Download button located in the bottom section 3 Once the file is downloaded it will be automa...

Страница 8: ...s website Service Office 48 22 51 88 320 or 48 22 51 88 322 Office hours are 9 00 a m to 5 00 p m Central European Time E mail at office svantek com Internet at www svantek com Address SVANTEK Sp z o...

Страница 9: ...nding Band Limiting filters Wf for motion sickness filter measurements according to ISO 2631 1 option RMS RMQ Detectors Digital true RMS RMQ detectors with Peak detection resolution 0 1 dB Measurement...