Connect the digital tester to the 12V and GND-1 check pins and adjust VR4 so that the voltage is 12.0V.
Connect the digital tester to the 5V and GND-1 check pins and adjust VR5 so that the voltage is 5.0V.
Connect the oscilloscope to the FRQ and GND-P check pins and adjust the waveforms at VR2 so that the below
is shown.
The value of frequency counter is 78.43 kHz.
12.75μsec(78.43 kHz)
-29-
Adjust the Power Supply PCB (No.654) for S-1800M10
(Reckless adjustment may cause serious problem.)