BV-10 OPERATION MANUAL – REV. 3.00 (20180423) 

21 

current-voltage relationship of a micropipette is non-linear; doubling the current may 

produce much more than twice the voltage drop. Changing the sign of the current can 
dramatically change the magnitude of the voltage drop. As a result, there is no single number 

that uniquely defines the resistance of a pipette. Thus, two accurate instruments employing 
different measuring currents could produce completely different measured values of an 

electrode's resistance and both would be valid. It follows that an electrode resistance meter 
need only have absolute accuracy to within a few percent to accommodate most users. 

However, it is important to measure resistance with high resolution, particularly when 
screening electrodes or when beveling. 
 

Содержание BV-10

Страница 1: ...One Digital Drive Novato CA 94949 Voice 415 883 0128 Web www sutter com Fax 415 883 0572 Email info sutter com BV 10 Micropipette Beveler Operation Manual Rev 3 00 20180423...

Страница 2: ...BV 10 OPERATION MANUAL REV 3 00 20180423 2 Copyright 2018 Sutter Instrument Company All Rights Reserved...

Страница 3: ......

Страница 4: ......

Страница 5: ...n for which it is designed SAFETY WARNINGS AND PRECAUTIONS Electrical Operate the BV 10 using 110 120 V AC 60 Hz or 220 240 V AC 50 Hz line voltage This instrument is designed for connection to a stan...

Страница 6: ...this device Operate this unit using the indicated line voltage This unit is designed for operation in a laboratory environment Pollution Degree I This unit is designed for connection to a standard lab...

Страница 7: ...ion of Package Contents 13 2 2 Set Up and Assembly Instructions 14 3 OPERATION 17 3 1 Grinding Procedure 17 3 1 1 General Information 17 3 1 2 Control of Beveling by Measuring Electrical Resistance 18...

Страница 8: ...1 1 Manipulator 9 Figure 1 2 Pedestal Grinding Assembly 10 Figure 1 3 BVM CE Electrode Resistance Meter 11 Figure 2 1 BV 10 beveler component layout 13 Figure 2 2 Pedestal grinding plate assembly 15 F...

Страница 9: ...ate rotates on a stationary optically flat pedestal by virtue of a magnetic coupling with a rotating drive ring Rotational speed is approximately 60 rpm Pipette positioning relative to the grinding pl...

Страница 10: ...Electrode Impedance Meter and BV 10S 40X Stereo Microscope 1 4 Technical Support Unlimited technical support is provided by Sutter Instrument Company at no charge to our customers Our technical suppor...

Страница 11: ...ontrol knob used to position the pipette just above the grinding plate surface a fine control knob micrometer used to control the actual beveling of pipettes The sensitivity of the fine control can be...

Страница 12: ...quartz surface a magnetic drive ring that rotates around the pedestal and upper and lower retaining rings which hold the grinding plate The grinding plate rotates wobble free on the flat quartz surfa...

Страница 13: ...ler Resistance is indicated by a precision analog meter which allows the user to easily identify changes in microelectrode resistance while beveling Each meter is calibrated using the active lead that...

Страница 14: ...To aid viewing the beveling process an optical attachment is available The unit consists of a 40x stereomicroscope and mounting hardware For large diameter tips 1 m the attachment allows the user to b...

Страница 15: ...x 1 1 Beveler motor base with motor housing halogen lamp and pedestal assembly 2 Reference wick 3 Reference wick holder 4 Pedestal oil 5 Degreaser 6 Operating Manual 7 Abrasive grinding plates 2 and r...

Страница 16: ...05 to 1 m and a 0 3 m alumina abrasive plate for beveling pipette tips in the 0 5 to 1 m range Diamond coated plates are available for beveling larger tips and can be ordered from Sutter Instrument Co...

Страница 17: ...hole in the mounting block The end of the bar with the threaded hole and flat face is directed to the left The end of the bar with the setscrew is directed to the right Orient the bar such that the f...

Страница 18: ...te the scope counter clockwise relative to the vertical post To adjust the angle of the scope relative to the beveler plate hold the scope with one hand while loosening the thumbscrew that clamps the...

Страница 19: ...e abrasive surface and then monitoring the progress of beveling until a predetermined criterion is met In cases where the tips are so fine that they cannot be viewed with a microscope monitoring the m...

Страница 20: ...er The electrode will first make contact with the saline ring and then with the grinding plate As soon as the first change in resistance is noted stop advancing the electrode If the resistance remains...

Страница 21: ...s from the tip Now continue to lower the tip until the character of the bubble stream changes usually it appears that the bubble stream has disappeared though it has probably broken up into a stream o...

Страница 22: ...ement from mechanical shocks When you wish to make a reading switch to the X5 range If the reading is less than 100 Mohms switch to the X0 1 range This procedure will give the greatest resolution When...

Страница 23: ...single number that uniquely defines the resistance of a pipette Thus two accurate instruments employing different measuring currents could produce completely different measured values of an electrode...

Страница 24: ...BV 10 OPERATION MANUAL REV 3 00 20180423 22 This page intentionally left blank...

Страница 25: ...with a Kimwipe1 Rinse the plate with distilled water and blot dry with a Kimwipe It is necessary to have the abrasive surface cleaned of grease and oil so that a thin layer of saline or water may be...

Страница 26: ...BV 10 OPERATION MANUAL REV 3 00 20180423 24 This page intentionally left blank...

Страница 27: ...sive plate for scoring If either surface is scored call Sutter Instrument for technical support 5 3 Beveling Takes a Long Time Abrasive rating of plate is too fine for desired tip size Try a larger ab...

Страница 28: ...BV 10 OPERATION MANUAL REV 3 00 20180423 26 This page intentionally left blank...

Страница 29: ...at the factory The cost of shipment both ways is paid for by Sutter Instrument during the first three months this warranty is in effect after which the cost is the responsibility of the customer The...

Страница 30: ...BV 10 OPERATION MANUAL REV 3 00 20180423 28 This page intentionally left blank...

Страница 31: ...0 to 20 m tip sizes 104E Diamond abrasive plate very fine 0 5 m light brown for 0 7 to 2 0 m tip sizes 104F Diamond abrasive plate extra fine 0 1 m green for 0 2 to 1 0 m tip sizes 007 Degreaser bott...

Страница 32: ...BV 10 OPERATION MANUAL REV 3 00 20180423 30 This page intentionally left blank...

Страница 33: ...wer entry module Figure C 1 Power entry module The fuse that is readily visible in the fuse holder when you take it out is the one that is active when the holder is installed A spare fuse is also stor...

Страница 34: ...ace only with the same type and rating Miniature Fuse 5 x 20 mm IEC 60127 2 250VAC Time Lag T Low Breaking Capacity Glass Tube At 110 120 VAC Rating 0 2A 200mA Examples Schurter 0034 3110 At 220 240 V...

Страница 35: ...g on abrasive plate used Grinding Surface Variation Less than 1 0 m Grinding Speed 60 RPM Beveling Angle Range 5 90 degrees adjustable Micromanipulator Coarse drive 0 075 in dial revolution Fine drive...

Страница 36: ...BV 10 OPERATION MANUAL REV 3 00 20180423 34 This page intentionally left blank...

Страница 37: ...eling by monitoring bubble size19 grinding procedure general information 17 Grinding Procedure 17 I Introduction 7 M mains fuses 3 32 Manipulator 9 models 8 Motorbase 8 N notes user 36 O operation pre...

Страница 38: ...BV 10 OPERATION MANUAL REV 3 00 20180423 36 meter does not work 25 no beveling occurs 25 pipettes break during beveling 25 Troubleshooting 25 U Unpacking and setting up 13 W warranty 27 NOTES...

Отзывы: