6. Installation
6. Installation
Before you install the sensor, make sure that all components listed
below are included in your package.
Qty Description
Item No.
1
S430 Sensor
S695 4300/S695 4302
1
Sealing ring
No P/N
1
Alignment key
No P/N
2
M12 plug
C219 0059
1
Instruction manual
No P/N
1
Calibration certificate
No P/N
6.1 Installation Requirements
To install the sensor a ball valve or a nozzle is needed.
•
The inner thread has to be G 1/2”.
•
The diameter of the hole must be
≥ 13mm, otherwise the shaft
can not be inserted.
8.2 Installation Procedure
6.2 Determination of the installation point
In order to maintain the accuracy stated in the technical data, the
sensor must be insert in the centre of a straight pipe section with
unhindered flow characteristics.
Remark
If the pipe diameter is higher than the maximum shaft length, it is
possible to install the sensor out of centre of the pipe. For this please
contact the manufacturer.
Unhindered flow characteristics are achieved if the section in front of
the sensor (inlet) and behind the sensor (outlet) are sufficiently long,
absolutely straight and free of obstructions such as edges, seams,
curves etc..
Please consider that enough space exists at your site for a adequate
installation as described in this manual.
13