Annual Inspection
41
EN
(for 5 seconds) and the display depicts
Overpressure
.
4. Reduce the pressure.
The warning ends when the pressure falls below 19 mmHg (nominal pressure
plus 4 mmHg).
10.5
Max. Device Pressure Test
1. Select the
Standard Insufflation Mode
as the operating mode.
2. Select the max. nominal gas flow.
3. Connect a manometer
(1)
and an open Veress cannula
(3)
to the insufflation
tube connection
(2)
.
4. Select the max. gas flow.
5. Start insufflation:
Press the
START
key.
The manometer registers a pulsing pressure increase. When the pressure sta-
bilizes, the manometer indicates a maximum pressure between 50 and 75
mmHg.
6. Stop insufflation:
Press the
STOP
key.
10.6
Gas Flow Rate Test
Test setup with open connection, without connected insufflation tube.
1. Adjust in the user menu the gas flow rate 2 (20 liters).
2. Press the
RESET
key (0.0 L must be depicted).
3. Start insufflation: Press the
START
key. Now begin timing for one minute.
4. Stop the insufflation after one minute. Press the
STOP
key.
The gas consumption should be between 15.5 - 20.5 liters.