Care and Maintenance
37
EN
individuals to perform repairs, adjustments, or alterations on the device or acces-
sories/peripherals.
Certification
Ask the service technician for a certificate after he or she has inspected the unit
or performed any service tasks. This certificate lists the type and scope of the ser-
vice as well as the date and name of the servicing company together with the sig-
nature of the service technician.
9.4
Resetting the Fuse (only for 2 pole breaker)
The fuse may have to be reset if:
• displays and LEDs (if available on your equipment) do not light up,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the power supply input
and to a safety socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does
not
have to be opened to reset the fuse.
1. Switch device off.
2. Disconnect device from power supply.
3. Remove power connection cable from mains socket.
4. The fuse switch is located next to the mains socket.
5. Check the switch setting of the fuse as depicted in Fig. 9-1.
6. Press the fuse switch to reset the fuse.
7. Use the power cable to reconnect the shockproof safety socket with the rear
socket.
9.5
Replacing the Fuse (only for inlets with fuse drawer)
WARNING!
Replace the fuse
Replace the fuse only with a fuse provided by the manufacturer (see chapter 14
"Accessories").
The fuse may be defective and in need of replacement if:
• Indicators and display (if present on the device) do not light up,
• the device does not function.
Check whether
• the main power supply cable is properly connected to the power supply input
and to a grounded safety wall socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does
not
have to be opened to replace the fuse.
Fig. 9-1
Fuse for 2 pole breaker
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...