Using and Controlling the AirSeal® iFS in its Different Modes
EN
to a second cannula and open the stopcock.
NOTE!
For the safety of the patient, it is advised to start intra-abdominal insufflation
with a Veress needle and at the lowest flow rate (level 1, key
(1)
). Exchange the
Veress needle with a conventional cannula once the nominal pressure is reached
and the initial insufflation phase ends.
5. Stop insufflation by pressing the
STOP
key.
When working in the
Smoke Evacuation Mode
, it is possible that body fluids are
extracted from the surgical field over the evacuation line. In order to prevent con-
tamination of the device, these are retained by the fluid trap of the Smoke Evac-
uation tube set filter component. The capacity of the fluid trap is limited. A
warning message and an audible signal will be emitted if the fill level
LOW
is
reached. Cannula and tubing placement should be checked to make sure no more
fluid enters the filter. At this point, change the filtered tube set. Insufflation can
be continued with full functionality. If fluid level
HIGH
is reached, the Smoke
Evacuation function will shut down but insufflation is still active.
If the
Smoke Evacuation Mode
has been stopped because the fluid level
HIGH
was reached, the iFS should be inspected to assess any possible damage. Call for
service.
6.3
Standard Insufflation Mode
The
Standard Insufflation Mode is intended for use with a Single Lumen Filtered
Tube Set and conventional cannulas.
1. Switch device on (consult section 5.5 "Switch Device On" for additional infor-
mation).
2. Press the
Standard Insufflation Mode
key to select this operating mode.
3. Connect the Single Lumen Filtered Tube Set.
Inserting the Single Lumen Filtered Tube Set
To be carried out by non-sterile technician:
• Open packaging of the Single Lumen Filtered Tube Set.
• Have a sterile technician remove the tube set from the inside of the pack-
age.
To be carried out by sterile technician:
• Keep the distal tube Luer lock connector in the sterile area and hand the
proximal tube end with filter to the non-sterile technician.
• Connect the Luer lock connector with the conventional cannula. Open in-
flow valve.
To be carried out by non-sterile technician:
• Attach the insufflation tube set to the insufflation tube set connection at
the front of the device (barbed connector).
4. Choose desired settings.
Setting the nominal flow and nominal pressure (increase/decrease) is possi-
distal end
proximal end
(3)
(1)
(4)(5)
(8)
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...