14
Initial Device Startup
EN
nection or connect to centralized CO
2
gas supply.
4.2.1
Connecting a Gas Cylinder
CAUTION!
Always use a high-pressure hose to connect gas cylinder and device.
The gas cylinder must be in a vertical position. The gas bottle pressure may not
exceed 80 bar/1160 psi. For the device start, the gas cylinder pressure must be at
least 25 bar/363 psi.
CAUTION!
Gas cylinders with riser pipes can release dirt and oily fluids into the device. Do
not use gas cylinders with riser pipes.
High pressure hoses
The following high pressure hoses are available:
Installation
High Pressure Hoses with PIN Connector
• Connect or loosen to the device using the open-end wrench SW 14.
• Connect or loosen to the gas bottle with your hand.
High Pressure Hoses with DIN Connector
• Connect or loosen to the device using the open-end wrench SW 14.
• Connect or loosen to the gas bottle using the open-end wrench SW 30.
High Pressure Hoses with ISO Connector
• Connect or loosen to the device using the open-end wrench SW 14.
• Connect or loosen to the gas bottle using the open-end wrench SW 32.
4.2.2
Connecting to Central Gas Supply
Use the following device connectors and high-pressure hoses available as addi-
tional equipment to connect to a central gas supply (house supply):
• Connector and hose for house gas supply NIST or
• Connector and hose for house gas supply DISS
Designation
High Pressure Hose Device US / Bottle DIN
High Pressure Hose Device US / Bottle ISO
High Pressure Hose Device US / Bottle PIN Index
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...