Chapter 10 Setting
www.supmea.com
21
This will reset all of the calibrated
parameters to defaul
10.18 Parameters Setting
Temperature Coefficient of solution: the
range is form 0.01 to 40.00%
Reference Temperature: the range is
from 15.0 to 35.0
℃
Note:In general, the conductance is
compensated back to 25.0
℃
, if the user
wants to compensate back to other
temperatures, the reference
temperature can be changed
10.19 Calibration
Put the electrode to the standard
solution.
It will display the EC reading.
When the EC reading is stable then
press UP/DOWN key to modify the cell
constant to make the EC reading
correspond to the standard solution.
If the EC Reading is over the EC range
or the temperature is over 0.0-60.0
℃
then it will display error message on the
button of LCD.