Chapter 10 Setting
www.supmea.com
20
10.16 EC Calibration
Press MODE key to enter the password menu and then press UP/DOWN/SHIFT
key to input password 1008 then press ENTER will enter to calibration mode or
press MODE key to exit. If no key is be pressed and over 10 minutes then it will
go back to measurement mode.
10.17 Menu
Press UP/DOWN key to select the
functions and then press ENTER key to
enter.
parameters setting: Input the
parameters.
Calibration: Calibrate the EC.
Reset parameters: reset all of the
calibrated parameters to default.
Notice:
If the cell constant is over than ±30%,
user should replace the new electrode.
10.18 Reset parameters