OPERATION
NOTE: Do NOT hold the button down while using the instrument, as was
necessary with the previous versions. Please read all instructions below.
To use the system, plug the headphones into the jack located on the right side
of the instrument. Slide the sensitivity adjustment all the way forward, towards
the nose of the sensor. The AccuTrak
®
VPE is now set at full sensitivity. To
turn on the instrument press the "ON/OFF" button for about one second - until
the display shows activity and you hear a beep in the headphones
–
and then
release the button. Do NOT hold the button down while using the
instrument, as was necessary with the previous versions. Always start at full
sensitivity, then if necessary, gradually reduce the sensitivity to reduce
background noise. Pressing the “ON/OFF” button again momentarily will
toggle on or off the “Peak Hold” function of the display, together with a
confirmation beep in the headphones. Peak Hold can be useful to better
observe the maximum strength signals being detected, and help find Leaks
and Faults. To turn off the instrument, press and hold the “ON/OFF” button
for at least two seconds, until it goes off. When using the instrument, if you
hear a high pitch tone and see the LED meter going to maximum this is the
“Over Range Audio Alarm” –
you need to reduce the sensitivity to a lower
setting if you want to properly interpret these stronger ultrasonic signals.
The AccuTrak
®
VPE is capable of hearing leaks from over forty feet away,
but it is best to hold the instrument as close to the test area as safely possible.
Check around fittings and piping using a zigzag motion, carefully covering all
suspected areas. If there is no ultrasonic sound present, the display should
not have any lights on. When an ultrasonic signal generated from a leak is